Instruction for the MRC-600 laser scanning confocal microscope
Note: A typical workflow can be found
at the end of this manual.
Starting the system
- Sign in
- check the log for unusual comments and sign out time of the
previous user
- Mercury Bulb (Optional, applicable only if you use conventional
epifluorescence to locate your specimen)
- you should only turn the power on if it is turned off at least 20
min ago
- check the counter of the power supply, if it is over 300 h please
notify Facility staff first before turning it on
- turn on the mercury lamp power supply: hit the power switch, press
and hold the IGNITE button for 1 to 2 s
Note:The READY light will come on very briefly, go out,
and then come back on again after the lamp is stable. The mercury
lamp should burn for at least 15 min. after ignition. It should not
be switched on and off unnecessarily. Please call the next user at
the end of your session to see if the mercury lamp can be left
on.
Warning: If the Mercury Lamp bursts, leave the room
immediately and notify Facility staff.
- Laser
- check to make sure that the fan is not running (which indicates the
laser has just been turn off), wait until it is off before switching
on the laser
- under the AV Table, turn the key on the power supply clockwise to
the horizontal position to switch on the laser
- the switch on the laser housing should be on "norm"—pointing
to the back
Note: The laser should not be switched on and off
unnecessarily. You can leave it running if it is going to be used
within the next 2 hr. Please call the next user at the end of your
session to see if the laser can be left on.
- Excitation wavelength
Tip: Most users will proceed to optimize the spatial
filters, to use K1+K2 one should select position # 3 (568) for
excitation filter, and # 3 (1 %) for neutral density filter now.
- rotate the Excitation Filter wheel in front of the laser housing to
select the desirable wavelength
Table 1 Excitation filters
| Position # |
Wavelength (nm) |
| 0 |
488 + 568 + 647, no filter |
| 1 |
488 |
| 2 |
488 + 568 |
| 3 |
568 |
| 4 |
647 |
| 5 |
488 + 568 + 647, no filter |
- Laser intensity
- rotate the adjacent Neutral Density Filter wheel to select the
desirable laser intensity
- the neutral density filters are the primary way to regulate the
amount of laser intensity impinging on your sample: start with #3
for fixed and #0 for live samples
- always use the lowest possible laser intensity (the least
transmission), increased laser intensity results in faster
bleaching of your sample
Table 2 ND filters
| Position # |
% Transmission |
| 1 |
10.0 |
| 2 |
3.0 |
| 3 |
1.0 |
| 0 |
0.1 |
- Scan Head
- turn the scan head on with the toggle switch on the back of the
unit, at the top and under the power cable
Front panel of the scan head

- GAIN should be mid-range (around 6) and AUTO is off (push-button
out)
- BLACK LEVEL should be on 5 and AUTO is off (push-button out)
- ENHANCE should be on LOW SIG. and External is off (push-button
out)
- select the Emission Filter
Table 3. Filter block selection
| Mode |
Channel 1 |
Channel 2 |
for Excitation (nm) |
| single |
BHS |
Blank |
488, position 1 |
|
YHS |
Blank |
568, position 2 |
|
RHS |
Blank |
647, position 3 |
|
Reflectance |
Blank |
488, position 1 |
| dual |
K1 |
K2 |
488 + 568, position 3 |
|
Reflectance |
BHS, YHS, or RHS |
use an excitation filter appropriate for channel 2 |
Note: see appendix for filter
specifications and the procedure for removal
and replacement of filter blocks.
Computer
- turn on both monitors
- turn on the computer if it is off, hit OK when prompted for windows
network login, no password needed
- double-click the desktop icon of CoMOS to launch the software
- if there is any initialization error, STOP and seek
help from Facility staff
Confirm the presence of the laser
- swing the perspex prism into the light path with the flat face outward
so the black target ring should be visible
- clear the light path by GENTLY swing the binocular head to the left,
and disengaging (pulling out) both epifluorescent filters
- start the laser scanning by pushing the spacebar on the keyboard, or
clicking the laser scan button in CoMOS
Caution: If there is no laser, check every component in the
light path; still nothing, STOP and notify Facility
staff.
- check to see if the laser is centered on the target, If not, try gently
rotate the filter block in Channel 1 to make sure it is properly seated;
if still not centered, STOP and notify Facility staff
Note: We do not adjust M1 anymore, this section in the appendix explains why.
- if the laser is flickering or throbbing (vs. gentle pulsating for a
scanning laser), it is off focus, please notify Facility staff for
adjustment.
Nikon Optiphot Microscope
- Objective lenses
- for optimal images, objective lenses should be clean
Note: The mixed pollen slide is good for checking the
dry objectives. If you see oil on the 20× or lower, please notify
Facility staff for cleaning.
- read the section below on using oil immersion objectives if you
will use them
- only use the supplied Ross lens paper; under no circumstance should
you need to rub the lens dry, not even with lens paper
- do not sweep dry lenses through immersion oil
- never use slides with wet nail polish or other media that may
damage the lenses
- Using oil immersion objective lens
- locate the object at a lower magnification, if needed
- lower the stage using the coarse focus knob (counterclockwise)
Caution: Turn the coarse focus knob SLOWLY and make
sure the objective is moving away from your slide. This help to
protect the objectives and your specimen.
- apply a drop of immersion oil on the coverglass
- rotate the objective assembly to bring the desirable lens into
position; use the turret, avoid grabbing the objectives
- raise the stage until the objective lens contacts the oil
- use the fine focus knob to bring your object into focus
- when you are done, carefully lower the stage
- swing the high power objective out and then remove your slide
- at the end of the session, take a piece of Ross lens paper and
gently draw it flat across the front lens surface in ONE direction
only
Caution: do not rub the surface in a circular or back
and forth manner
- 3 to 4 passes should remove most of the oil on the objective
lens
- Transmitted light (bright field)
- Bright field
- use the thumb wheel in the front to turn on the transmitted
light
- select A on the condenser turret
- locate your specimen first using low power (e.g., <
10×)
- adjust for Kohler illumination (only needed if you will acquire
images with the transmitted light detector or you need optimal
conditions to locate your sample)
- select the objective that you will use for the
experiment
- focus on the specimen
- fully open the Aperture Diaphragm and rotate the condenser
to
- A for brightfield
- appropriate position for phase contrast
- close down the Field Diaphragm with the knurled plastic
ring on the base
- focus on the leaves of the iris diaphragm with the
Condenser Focus Knob
- open the Field Diaphragm until the leaves are just within
the field of view
- center the iris diaphragm with the 2 Condenser Centering
Screws
- close down the Aperture Diaphragm until the intensity just
start to drop
Note: This reduces glare and corresponds to
about 70 % of the field.
- Phase contrast (not covered here)
- Epifluorescence
- gently push the appropriate lever in to engage one of the 2 filter
blocks
- B-2A: EX 450–490, DM 510, BA 520
- G-1B: EX 541–551, DM 565, BA 590
- open the epifluorescence Shutter only when you are examining your
specimen
Optimize the spatial filter
The spatial filters: M4 and M5 are mirrors directing signal to each of
the 2 detectors. Each filter has 2 Allen socket adjustment screws. Failing
to properly align the spatial filters will lead to degraded images.
Adjusting M4, the spatial filter for PMT1
Note:
- This procedure is for single channel or Channel 1 of dual channel
imaging.
- For single channel, use the green plastic slide for 488 nm, red one
for 568 nm, and the tetraspeck preparation for 647 nm.
- For dual channel imaging with K1+K2, use 568 nm and the red plastic
slide.
- with transmitted light at 4×, focus on an edge of the cross-mark on
the surface of the plastic slide or the tetraspeck beads
- increase magnification to 20× and focus
- turn off the transmitted light and prepare for laser scanning (see step
2 in Confirm the presence of the laser)
- select the appropriate excitation filter (see note above)
- a good starting point is: ND 3, GAIN 6, BLACK LEVEL 5 for Channel 1,
aperture at half open
- select PMT1 and start scanning
- focus on an edge of the cross-mark
- If you can't find the cross-mark, check the light path; still not, STOP
and get help from Facility staff
- move over to an unmarked area
- apply SETCOL.LUT (see Selecting LUTs)
- adjust the BLACK LEVEL properly e.g., the black corners on the right
should just turn solid green
- close aperture down to the minimum, adjust the GAIN to obtain some red
pixels such that any slight increase or decrease in signal can be
detected easily
- use the fine focus to get the brightest possible image which is the
surface of the slide, you should see a "hot spot" somewhere on the
monitor
- adjust the 2 screws for M4 in turns to center the "hot spot", adjust
the GAIN as needed
Note: M4 is very sensitive, so only make very slight turns
(< 1/16 of a turn)
Adjusting M5, the spatial filter for PMT2
Note:
- This procedure is for Channel 2 of dual channel imaging and you MUST
adjust M4 first.
- For K1+K2, use 488 nm and the green plastic slide.
- For reflectance use the green plastic slide for 488 nm, red one for
568 nm, and the tetraspeck preparation for 647 nm.
- do the same as in Adjusting M4, except use Channel 2
and select PMT2
- with PMT2, you can expect to get an evenly illuminated field when M5 is
properly adjusted
- M5 is not as sensitive as M4 so larger displacement (1/4 turn) is
needed to produce any noticible signal change
Using CoMOS
Scanning
- select the correct detector (PMT1, PMT2, BOTH) on the left
- click on the top left large laser icon or press the space bar to toggle
on or off laser scanning
Selecting Lookup table (LUT)
- go to Display, select Output LUTs ...
- click Open at the bottom of the Edit Output LUTs dialog box
- in the Load Output LUT File dialog box, click on the desirable file
e.g., SETCOL.LUT, and then OK to load the LUT
- go to Display and toggle on "Hold LUTs" (a check mark will appear
before it)
- to return to the default greyscale LUT, either toggle off "Hold LUTs"
and scan, or select Monochrome in the Edit Output LUTs dialog box
Collecting a Z-series
- locate your object of interest, focus, and adjust Zoom, GAIN, BLACK
LEVEL, and aperture for an optimal image
- set Objective to the same as used on the microscope
- select ON for the focus motor
Note: Whenever you turn the focus motor on, the current
position of the stage is defined as 0.
- select an appropriate Z-Step size
- while scanning, focus to the top of the region you wish to image by
clicking the left arrow for Position which moves the stage down
Note: While scanning through the whole volume, you should
also watch out for over- or under-exposed pixels in the area of
interest, adjust gain and black level as needed.
- click the Z-Stop button to set the upper limit of the scan, or you can
type in the number directly
- using the right arrow for Position, focus through your sample (the
stage moves up) to find the lower limit of the scan
- click the Z-Start button to set this limit
- select a Collection Filter and set the number of passes (N scans) and
Factor, as needed
- from Collect, select Z-Series\Time Series… to bring up the Parameters
dialog box
- make sure that "Save to File" and "Enable Z-Step" are checked and click
OK
- in the Save Z-Series to File dialog box, enter a file name (8
characters max), make sure Byte (8-bit) is selected, and click OK to
begin collecting your data
Note:Files are saved in C:\COMOS by default, if you want to
save your file elsewhere, enter the relative path for the first file
and all subsequent file in the session will be saved in that
subdirectory. E.g., you want the files to go into
C:\COMOS\IMAGES, create the subdirectory in windows first, then for
your first save, type in IMAGES\filename
Note: Your files can be stored temporarily on the harddrive
for up to a month. However,files will be erased sooner if disk space is
low. Please back up your files and make sure they are readable as soon
as possible.
Some tips on obtaining an image
- at a Scan Speed of F2 each scan takes 1 s, you should adjust the
settings slowly while scanning so you can see the changes
- if you do not know the response of your sample
- select a low laser power e.g., ND position 3 for fixed and 0 for
live samples
- open the aperture completely
- set the GAIN to about 8
- leave the BLACK LEVEL at 5
- start scanning
- if there is nothing, stop scanning and check the light path
- if everything is set correctly, start scanning and then increase
the GAIN slowly, adjust the focus as needed
- if still nothing with GAIN at 10.0, select the next higher ND
filter
- unless your sample is really weak or completely out of focus, you
should start to see something
- as your object comes into view, you can start to optimize it by
checking the focus, and then picking an appropriate setting for the
aperture, GAIN, ND filter, and BLACK LEVEL
Image Optimization
- to maximize the dynamic range, apply SETCOL.LUT to
help set the correct gain and black level first and then fine tune them
for your image (see the appendix for an explanation
about using SETCOL)
- use as large an aperture size as you can tolerate so you can use less
laser intensity and lower gain
- use Kalman: select the number of passes in N scans
- use a slower scan speed—more laser exposure!
- for really low signal, try the accumulate filters
Routine Shutdown (as listed in the log
sheet)
- select OFF for the focus motor, exit CoMOS
- turn off the laser and scan head
- start transferring your files using the zip, or network
- if the Hg bulb is still on, call the next user to ask if the Hg bulb
can be left on
- if you finish early, notify the next user
- make sure the transmitted light is off
- remove any excess oil from the 60× objective
- rotate the objective turret to put the prism in the light path
- turn off the image monitor
- when you are done transferring your files, turn off the menu
monitor
- put the cover on the microscope
- complete the log and record Time Out
- cleanup the work area before you leave
Appendix
Setting the gain and black level (offset)
At each pixel of the specimen, the fluorescent photons are collected and
amplified by the photomultiplier tube. The analog output signal (voltage) is
then passed to an analog to digital converter (ADC). The 8-bit ADC assigns
256 levels to represent the image. The value 0 is assigned as the minimum
intensity whereas 255 is the maximum intensity. By setting the black level,
you tell the ADC that below a certain voltage everything should be assigned
as 0; this controls the brightness. The gain, on the other hand, defines the
range and thus sets the highest voltage beyond which the ADC will assign as
255; it determines the contrast.
In order to display the image on the monitor, you instruct the computer to
map the 256 levels to some display characteristics according to a LookUp
table (LUT), e.g., to display a grey scale image, 0 is choose as black, 255
white, and shades of grey in between. To make full use of the ADC’s
capability, you will need to make sure that the black level and gain are
selected properly for the parts of the image you are interested in. Since
human eyes are less sensitive to variations at the 2 extreme ends of the
range, mapping these display levels to a different color can help ascertain
these settings, e.g., SETCOL.LUT maps 0–5 to green and 250–255 to red.
The specification of the filter blocks
Table 4. Filter blocks
| Mode |
Channel 1 |
Channel 2 |
for Excitation (nm) |
| single |
BHS
Dichroic: 510 LP*
Emission: OG 515 LP
|
Blank |
488 DF10, position 1 |
|
YHS
Dichroic: 585 DRLP
Emission: 585 EFLP
|
Blank |
568 DF10, position 2 |
|
RHS
Dichroic: 660 DRLP
Emission: 680 EF32
|
Blank |
647 DF10, position 3 |
|
Reflectance |
Blank |
488 DF10, position 1 |
| dual |
K1
Dual band dichroic
|
K2
Dichroic: 560 LP
Emission: 522 DF35
Emission: 588 EFLP
|
488 + 568, position 3 |
|
Reflectance |
BHS, YHS, or RHS |
use an excitation filter appropriate for channel 2 |
- *DF:
- discriminating filter with very steep-sided passbands
- DRLP:
- dichroic longpass filter
- EF:
- edge filter
- EFLP:
- longpass edge filter
- LP:
- longpass filter
- OG:
- orange glass (absorbs blue light)
Removal and Replacement of the Filter
Blocks
Caution: ALWAYS stop scanning first before changing filter blocks
and do NOT touch the optical surfaces.

Why you should not adjust the mirror, M1
It is critical that M1 be centered or you will experience a severe drop in
intensity and uneven illumination. It should only be adjusted by Facility
Staff with the REFL block installed. All other blocks should have the beam
centered by a slight and gentle rotation of the block.
Here is the reason. The filter block is held in place by 3 pins against
the mounting ring. Wear and tear on the pins and the seats on the mounting
ring has created some play which allow the block to rotate a few degrees upon
its longitudinal axis. Therefore, it is not always possible to place the
block in the appropriate position. The REFL block is the least worn and still
maintains the factory alignment. Once the beam is centered with the REFL
block, the correct placement of the other blocks on the mounting ring can
simply be found by keeping the beam in the center of the target.
A Typical Workflow
Caution: This is NOT a step by step instruction but an example
of a typical workflow. You should follow the specific instructions in the
respective sections in this manual.
- check log for unusual activities or errors
- sign in
- powerup laser
- powerup scan head
- launch CoMOS
- optimize M4 (and M5, if applicable)
- locate specimen at low magnification
- move to your area of interest
- begin scanning and get data!
- quit CoMOS
- poweroff laser
- poweroff scan head
- transfer data out
- sign out
- clean up