Instruction for the Radiance2000 MP laser scanning confocal
microscope
Note: A typical workflow can be found
at the end of this manual.
Starting the system
- Sign in
- Check the log and extended log for unusual comments
- Mercury Bulb (Optional, applicable only if you use conventional
epifluorescence)
- check the log, you should only turn the power on if it is turned
off at least 20 min. ago
- check the counter of the power supply, if it is over 300 hr. please
notify Facility staff first before turning it on
- turn on the mercury lamp power supply: hit the power switch, press
and hold the IGNITE button for 1 s to 2 s
Note:The READY light will come on very briefly, go out,
and then come back on again after the lamp is stable. The mercury
lamp should burn for at least 15 min. after ignition. It should not
be switched on and off unnecessarily. Please call the next user at
the end of your session to see if the mercury lamp can be left
on.
Warning: If the Mercury Lamp bursts, leave the room
immediately and notify Facility staff.
- Dell PowerEdge 1300 Computer
- if the computer is not already on, make sure the ICU is off before
turning on the computer
Caution: Do not turn on the ICU (lasers + PMTs) until
the computer has finished booting up, or you will damage the ICU.
Do not launch Lasersharp 2000 until the ICU is on or it will not
find the hardware components. If the computer needs to be
restarted, the ICU must be turned off first.
- if the computer is left on by the previous user, it is a good idea
to reboot the computer for your session
- make sure the ICU is off, then turn on the monitor and boot up the
computer
- press CTRL-ALT-DEL and log in as _____ (passwd: _____)
Note: If you wish to burn a CD, you may have to log in
as a different user. Please see burn a CD in Moving files.
- Instrument Control Unit (ICU) and lasers
- when the computer has finished booting, go behind the AV Table,
turn on the ICU with the rocker switch near the top (the key is
always in the ON position, horizontal)
- you should hear a series of snaps indicating mirrors and filters
are being moved into position
Note: If you hear a dull noise instead, the
initialization has probably failed and LaserSharp will not launch
properly.
- launch LaserSharp and log in
Note: If hardware check fails, quit LaserSharp, turn
off ICU, wait 1–2 min., and then restart ICU. Note this event in
the error log. If it fails again, please stop and notify Facility
staff.
- turn on only the laser you will need by pushing the round buttons
on the front panel
- Kr/Ar: 488 nm and 568 nm
Caution: The Kr/Ar laser will beep for about 30 s
during startup. The cooling fan for the Kr/Ar laser is
automatic and will continue running for a few minutes after the
laser is powered off; even if the ICU is switched off. Do not
turn on the laser while the fan is still running, doing so will
decrease the life span of the laser.
- Red diode: 637 nm
Nikon Eclipse E600 FN Microscope
- Objective lenses
- for optimal images, objective lenses should be clean
Note: If you see oil on the dry objectives 20× or
lower, please notify Facility staff for cleaning. The mixed pollen
slide is good for checking the dry objectives
- read the section below on using oil immersion objectives if you
will use them
- only use the supplied Ross lens paper; under no circumstance should
you need to rub the lens dry, not even with lens paper
- do not sweep dry lenses through immersion oil
- never use slides with wet nail polish or other media that may
damage the lenses
- Using oil immersion objective lens
- locate the object at a lower magnification, if needed
- raise the objective assembly using the coarse focus knob (clockwise)
Caution: This is a fix stage microscope. Turn the
coarse focus knob SLOWLY and make sure the objective is moving away
from your slide. This help to protect the objectives and your
specimen.
- apply a drop of immersion oil on the coverglass
- rotate the turret of the objective assembly to bring the desirable
lens into position; avoid grabbing the objectives
- lower the objective assembly SLOWLY onto the slide until the
objective contacts the oil
- use the fine focus knob to bring your object into focus
- when you are done, carefully raise the objective assembly
- swing the high power objective out and then remove your slide
- at the end of the session, take a piece of Ross lens paper and
gently draw it flat across the front lens surface in ONE direction
only
Caution: do not rub the surface in a circular or back
and forth manner
- 3 to 4 passes should remove most of the oil on the objective
lens
- Transmitted light (bright field and DIC)
- Bright field
- push the Light Path Selector to BINO
- gently slide the Cube Selector lever to INT
Caution: do not pull the lever out or it will
jam
- turn on the transmitted light with the controller next to the
monitor

- make sure the DIC Analyzer located in the center of the
Epifluorescence Module is out
- locate your specimen first using low power (e.g., <
10×)
- adjust for Kohler illumination (only needed if you will acquire
images with the transmitted light detector or you need optimal
conditions to locate your sample)
- select the objective that you will use for the
experiment
- focus on the specimen
- fully open the Aperture Diaphragm and rotate the condenser
to
- A for brightfield
- DIC L for DIC at 10×
- DIC M for DIC at 20×
- DIC H for DIC at 40× and 60×
- close down the Field Diaphragm with the knurled plastic
ring
- focus on the leaves of the iris diaphragm with the
Condenser Focus Knob
Caution: the condenser can travel through the
stage and pop out your slide if you raise it up too far.
- open the Field Diaphragm until the leaves are just within
the field of view
- center the iris diaphragm with the 2 Condenser Centering
Screws
- close down the Aperture Diaphragm until the intensity just
start to drop
Note: This reduces glare and corresponds to
about 70 % of the field.
- Differential Interference Contrast (DIC)
- do 1 to 3 as in Bright field above
- insert the DIC Analyzer located in the center of the
Epifluorescence Module
- do 5 as in bright field but leave the Aperture Diaphragm fully
open
- rotate the Substage Polarizer until the fine structures of your
specimen gain sufficient contrast
- Epifluorescence
- gently slide the Cube Selector lever to engage one of the 3 filter
blocks
- UV-2E/C: EX 330–380, DM 400, BA 435–485
- B-2E/C: EX 465–495, DM 505, BA 515–555
- G-1B: EX 541–551, DM 565, BA 590
Caution: do not pull the lever out or it will
jam
Note: If a specific block is not in the microscope,
please see Facility staff. There is additional information
about the filter blocks in the appendix.
- disengage the DIC Analyzer
Caution: the Analyzer can be pulled out completely and
fell off the microscope!
- open the epifluorescence Shutter only when you are examining your
specimen
Main features of LaserSharp2000
(see the appendix for Basics in LaserSharp2000)
- Acquire an image (assuming you have located an object of interest at
the desirable magnification)
- slide the Cube Selector lever to the appropriate position (do not
pull the lever out!)
- INT for the conventional PMT with all visible wavelength lasers
and Mai Tai
- DDS for the Direct Detector with the Mai Tai alone
- pull the Light path selector out to PHOTO
- pull out the DIC Analyzer (don't drop this!)
- make sure transmitted light is off, Epifluorescence Shutter is
close, and ICU is on
- double-click the LaserSharp2000 icon (not the No Hardware version)
and login
Note: The software will load the previously used method
and open a new experiment.
- in the upper most pane of the Control Panel: Microscope
- set a speed and zoom
- select the objective to match the one you are using
- select Direct
- set the resolution (Pixels × Lines) of your image
- in the middle pane of the Control Panel: Channels
- select either a Simultaneous or Sequential setting
- move the sliders on the Gain, Iris, and laser power to a
desirable value
Note: The Lock buttons will lock the parameters
across the Simultaneous and Sequential setting.
- start scanning by clicking the Exploration icon (topmost, far left
button)
- optimize the image by picking an appropriate setting for the Iris,
Gain, laser power, and Offset with respect to the nature of your sample
Note: There are compromises in any parameter changes.
- increase gain enhances both signal and noise; this can be partially compensated for using Kalman filter or slower scan speed, but with increase exposure to laser
- increase the confocal aperture size (Iris) allows more out of focus signal to come through to yield a brighter but less sharp image; if the aperture is wide open, you are essentially using the equipment as a glorified epi-fluorescence microscope.
- higher laser power increases signal but will bleach your preparation faster
- Some tips on obtaining an image
- rough in the settings with the slider, and then fine tune with the
small arrow keys; adjust the settings slowly while scanning your
sample so you can see the changes
- If you do not know the response of your sample
- pick a Sequential setting to start with
- select a low laser power e.g., 1.0
- open the Iris completely
- slide the Gain to about 30
- leave the Offset at 0.0
- start scanning by clicking the Exploration button (topmost, far
left button)
- if there is nothing, stop scanning (click Exploration) and
check the light path
- if everything is set correctly, start scanning and then pull up
the Gain slowly (as appropriate for the scan speed), adjust the
focus as needed
- if still nothing with Gain at 100, pull up the laser power
SLOWLY
- unless your sample is really weak or completely out of focus,
you should start to see something
- as your object comes into view, you can start to optimize it by
checking the focus, and then picking an appropriate setting for
the Iris, Gain, laser power, and Offset
- Image Optimization
- to maximize the dynamic range, apply SETCOL.LUT to help set the
correct Gain and Offset first and then fine tune them for your image
(see the appendix for an explanation about using
SETCOL)
- use Kalman: select the number of passes in the N box (reselect
Direct when you are done taking your image)
- start with an aperture size optimized for the objective
selected—Target Iris, click on the target icon beneath the slider,
and open it as wide as you can tolerate so you can use a lower gain
Note: Target Iris is the Airy disk diameter as
calculated with respect to the NA of the objective and the
refractive index of a common mountant.
- use a slower scan speed
- the Optics Diagram page (topmost, second right button) allows you
to change the following even as you are scanning
- Signal Enhancing Lens
- Dichroic Mirror
- Emission Filter
- Polarization Filter
- Blocking Filter
- Power Meter setting for the Mai Tai
Note: Occasionally, the software fails to perform a
mirror/filter change that you have selected. If you do not hear the
clicking, it is likely that it does not occur. Try again, and if
the problem persists, please notify Facility staff.
- Collecting Z-series data (XY Series; XZ Series, or Line Scan are
possible)
- the lowest pane of the Control Panel: Focus motor
- turn on the Focus motor, the current Position will be defined
as 0
Note: Whenever you turn the focus motor on, the
current position of the objective assembly is defined as 0.
- set a step size appropriate for the object of interest and the
objective used
- step through your object with the arrow keys in the Position
box and define the Start and Stop position of the series, and
also adjust for the optimal exposure of the image
Note: click on the Start button will set the
current position as the start of the series, or simply enter
the value in the box
- before collecting the Z-series, confirm that all of the
software settings are correct, especially if you want to do
Kalman averaging
- click on the XY Series icon to open the XY Images window
- click the Enabled box under the Focus Motor tab
- verify the values in the Position boxes and the details given
in Collection Information
- click Start to collect the images
- save the series when it's done, your files should only be in
F:\Radiance files
Caution: Files saved elsewhere will be deleted
without further notice.
Mai Tai infrared laser: 780 nm to 920 nm (effectively 390 nm to 460
nm)
Start up
- if the water chiller and T40 Power Supply is on, skip to step 4 below
Note: To maintain temperature equilibrium, both the water chiller and the T40 power supply should be left on at all times. In
emergencies, please turn off the T40 first and then the water chiller.
Follow the directions below to re-start the system.
- if the water chiller is off, switch it on
Caution: The chiller should be set to 21 °C; wait until
the temperature settles before switching on the Mai Tai’s power
source. THE WATER CHLLER MUST ALWAYS BE ON WHEN THE MAI TAI POWER
SUPPLY (T40) IS ON. LASER COMPONENTS WILL BE FATALLY DAMAGED
OTHERWISE.
- under the AV table, switch on the T40 power supply using the orange
rocker switch
- boot up the Dell Optiplex computer if it is off
- press CTRL-ALT-DEL and log in as _____ (passwd: ____ )
- double click the Mai Tai software icon
- turn on the Pockel’s cell power source
Note: The Pockel's cell modulates the laser power. If it is
not on, you will be imaging with full laser power!
Mai Tai Controller software basics
- once launched, the software will monitor the operation of the pumping
lasers as indicated by the flashing RS-232 icon
Note: if the T40 is just powered up, a “% warmed-up”
will be displayed. The laser should only be turned on when the System
Status reads “Ready to turn on”.
- select a wavelength around 800 nm for start (you can select any
wavelength later after the laser is stabilized)
- click and hold the ON/OFF button until “EMISSION” displays; the
Power Meter will rise, stabilize, and begin pulsing (box is green)
in a few minutes; if not, STOP
- select the desirable wavelength
Note: Move the gray slider to the desired wavelength with
the mouse and release or type the desired wavelength in the Set
Wavelength box (or use the arrow keys). You can rely on published
literature or vendor specification in selecting the wavelength, or
sometimes the easiest way to find the best wavelength to excite your
fluorophore is by observation. Tune across all the wavelengths while
scanning, and watch until the brightness of the fluorophore is the
strongest.
- if the Shutter button is white (close), click and hold the button until
it turns red (open)
- to power off the laser, click once on the ON/OFF button
Using the Direct Detector
Note: It is not possible to use the DDS and the other PMT in the
same method.
- select a method that uses DDS or create a new one, DD1 is the green and
DD2 is the blue channel
- open the Filter Carousel Wheel Cover and check the Filter Carousel
Wheel to make sure the desired cube is selected
Filter Carousel Wheel Elements
| position |
name |
emissionA (DD1) |
dichroic |
emissionB (DD2) |
| 1 |
empty |
empty |
empty |
empty |
| 2 |
B/G |
HQ 515/30 |
DC 500 LP |
empty |
| 3 |
empty |
empty |
empty |
empty |
| 4 |
495 DCXR |
HQ 528/50 |
495 DCXR |
HQ 450/80 |
- make sure the Filter Carousel Cover is properly installed (window is
closed)
- gently slide the Cube Selector lever to the DDS position (far left)
Moving Files
- via the network using SSH secure file transfer
- double click the SSH icon on the desktop
- select the profile "Imaging Facility" or other server you
prefer
- enter the password when prompted
- navigate to your folder or create a new one as needed
- open up Windows NT Explorer, navigate to your files, drag and drop
them into the right pane of the SSH window
- burn a CD
- simple
- double click the CD icon on the desktop
- either use the wizard or click return to RecordNow to make a new
data CD
- open up Windows NT Explorer, navigate to your files, drag and drop
them into the RecordNow window
- burn away!
- more options
- you need to log in as _____ (passwd: _____ )
- double click the create CD icon on the desktop and click data CD
- on the upper panes, navigate to your files, drag and drop them into
the lower right pane
- 2 options available in layout (File | CD Layout properties) are
worth notice
- You can limit the filetypes to write to the CD by going to the File type tab, select "do not add ..."
- If you are going to write an ISO 9660 CD, make sure the filenames are under 30 characters long. The Joliet file system does not have this limitation but Mac OS 9 or earlier will need extra software to read the CD. The file system for the CD is specified in the General tab.
- burn away!
- this is a privilaged user accout, please log out promptly when you are done
One way to save time and storage space: You can select just the raw PIC (or other files) to transfer or burn to CD. However, this will prevent you from opening your experiments ever again in LaserSharp. If you will never use LaserSharp to view these files again, this should not be a problem.
- it will make life a lot easier if you put all the files in the
current session in a single folder
- do a file search in windows, enter "
*.pic" in the box labelled, add ", *.tif" if you have any TIFF to move or add other file types separated by coma as needed
- browse to your folder and hit search
- select the files in the result window, drag and drop them into the
SSH or RecordNow window
Routine Shutdown (as listed in the log sheet)
- for Mai Tai only
- power off the Mai Tai, exit the software, and turn off the
monitor
- turn off the Pockel's cell
- check and make sure that the water chiller and T40 are still on
Note: If you need to shutdown the Mai Tai system, turn
off the T40 first and then the water chiller.
- complete the Mai Tai log and record the Time Out
- all lasers
- exit LaserSharp
- turn off the laser and ICU
- start transferring your files using the zip, CD or network
- if the Hg bulb is still on, call the next user to ask if the Hg
bulb can be left on
- if you finish early, notify the next user
- make sure the transmitted light is off
- remove any excess oil from the 40× or 60× objectives
- rotate the objective turret to put the prism in the light path
- when you are done transferring your files, shutdown the computer
and turn off the monitor
- put the cover on the microscope
- complete the Radiance 2000 log and record Time Out
- if applicable, provide details for any problems encountered in the
Radiance 2000 Extended log and notify Facility staff
- cleanup the work area before you leave
Appendix
Setting the gain and black level/offset
At each pixel of the specimen, the fluorescent photons are collected and
amplified by the photomultiplier tube. The analog output signal (voltage) is
then passed to an analog to digital converter (ADC). The 8-bit ADC assigns
256 levels to represent the image. The value 0 is assigned as the minimum
intensity whereas 255 is the maximum intensity. By setting the black level,
you tell the ADC that below a certain voltage everything should be assigned
as 0; this controls the brightness. The gain, on the other hand, defines the
range and thus sets the highest voltage beyond which the ADC will assign as
255; it determines the contrast.
In order to display the image on the monitor, you instruct the computer to
map the 256 levels to some display characteristics according to a LookUp
table (LUT), e.g., to display a grey scale image, 0 is choose as black, 255
white, and shades of grey in between. To make full use of the ADC’s
capability, you will need to make sure that the black level and gain are
selected properly for the parts of the image you are interested in. Since
human eyes are less sensitive to variations at the 2 extreme ends of the
range, mapping these display levels to a different color can help ascertain
these settings, e.g., SETCOL.LUT maps 0–5 to green and 250–255 to red.
Basics in LaserSharp2000
- The online help file and the Radiance 2000 MP manual
(c:\Radiance_pdfs\9M60UM04.PDF) provide some useful information even
though they are based on an older version of the software.
- The Experiment is a basic unit of operation in LaserSharp. It is a
collection of files in a folder containing the raw images from each
channel, the Merge Pane, and other derivatives. After collecting a XY or
XZ Series, or Line Scan, you must either save or discard the current
result before you can launch a new experiment to collect more images.
- The Experiment can only be opened in LaserSharp whereas the raw pics
can be opened in any software that can read BioRad PICT files e.g.,
ImageJ.
- If you change the name of any folders and files in the Experiment,
LaserSharp will fail to open the files; except in the case with a Merge
Pane, the merged images (xxx_MergeData.pmf) can be deleted to save disk
space without affecting the integrity of the Experiment. LaserSharp can
regenerate the merged image. However, you can import any BioRad PICT
files by right clicking an Experiment listed in the Experiment
Browser.
- Images can be exported at anytime when an experiment is opened by right
clicking on the image, select export, and choose an appropriate file
type.
- Right clicking on the image will also allow for various operations such
as projections, merge, orthogonal view, etc.
- The Method contains details for the Simultaneous and Sequential
settings. These include the optical configuration and most of the
acquisition parameters displayed in the Microscope and Channels Panes of
the Control Panel. Any changes made in the Optics Configuration and the
acquisition parameters can be saved by going to Methods | Save Settings.
The file Creating Methods (c:\Radiance_pdfs\9M60UM31.PDF) provides
examples in dealing with methods.
Note: when you create a new method, put your username in
parenthesis at the end of the name to make it unique. All methods on
this computer must have an unique name.
- Depending on the amount of RAM available, you can open any previously
acquired experiments. The file will be added to the Experiment Browser
window where you can simply click to open the experiment again.
- Your files can be stored temporarily in D:\Radiance Files for up to a
month. However, files will be erased sooner if disk space is low. Please
back up your files and make sure they are readable as soon as possible.
Caution: Files saved elsewhere on this computer will be
deleted without further notice.
The Epifluorescense Attachment Slider
Caution: Do not try this unless you have been trained on this
procedure!
There are labels in the front of the
Attachment which show the configuration of the Slider. The default is
(from left to right): DDS, INT, G-1B, B-2E
- Filter cubes can be accessed from either side of the Attachment by
opening or slide out the protective cover.
- The physical position of the cubes on the slider is (from left to
right): B-2E, G-1B, INT, DDS
- Filter cube on the far left or right can be removed to accommodate the
UV-2E cube: if you do not use DDS, take the far right one out; otherwise,
take the far left (B-2E) out.
- The necessary tools (2 mm Allen wrench) and filter cubes are in the
drawer opposite the microscope
- Procedure to replace the cube
Caution: make sure you do not touch any of the optical
surfaces!
- open or slide out the protective cover
- move the slider to one end to expose the appropriate filter
cube
- loosen up the set screw securing the cubes 1 to 2 turns
- grab the filter cube and slide it off the rail
Note: The big round lens is on the right side of the
DDS cube and there should be a filter on the top (barrier or
emission) and the back (excitation) of the other filter cubes.
- slide on the replacement cube and tighten the set screw to
finger-tight
Caution: Do not overtighten!
- replace the cover
- put or replace the label on the front in the appropriate slot to
indicate the change e.g., if you take out the DDS, put the label in
the far left slot
- put the replaced cube in one of the protective plastic box together
with its label
- restore the default configuration at the end of the session unless
the next user requested the same configuration
A Typical Workflow
Caution: This is NOT a step by step instruction but an example
of a typical workflow. You should follow the specific instructions in the
respective sections in this manual.
- check log for unusual activities or errors
- sign in
- reboot or startup computer
- start up ICU
- launch Lasersharp
- powerup laser
- locate specimen at low magnification
- move to your area of interest
- begin scanning and get data!
- quit lasersharp
- poweroff laser
- poweroff ICU
- transfer data out
- sign out
- clean up