O2 Gas Laboratory Comparison
O2 gas from a compressed cylinder originating from Praxair in Seattle, Washington.
Here is a table of all laboratories that have measured this gas, the results they obtained, and information provided about their reference materials. ± indicates standard deviation. 17Oexcess was calculated as (ln(δ17O/1000+1)-0.528*ln(δ18O/1000+1))*1000000 and are in units of permeg.
Laboratory Number | δ17O (permil) |
δ18O (permil) |
17Oexcess (permeg) |
Reference Gas δ17O (permil) |
Reference Gas δ18O (permil) |
| 1 | -4.52 ± 0.023 | -8.55 ± 0.052 | 4.24 | 23.70 | 46.13 |
| 2 | -4.45 ± 0.022 | -8.60 ± 0.017 | 100.51 | unk | unk |
| 3 | -4.32 ± 0.009 | -8.26 ± 0.015 | 48.38 | -7.02 | -13.38 |
| 4 | -4.21 ± 0.068 | -8.08 ± 0.092 | 66.06 | 12.53 | 24.56 |
| 5 | -4.17 ± 0.044 | -7.89 ± 0.025 | 7.20 | 8.84 | 17.56 |