An Introduction
to Mass Spectrometry (cont.)
II. Mass
Analyzer & Ion Detector
Once ions are created, individual mass-to-charge ratios (m/z) are separated
by a second device, a mass analyzer, and transferred to the third device,
an ion detector. A mass analyzer uses some physical property (e.g., electric
or magnetic fields or time of flight) to separate ions of a particular
m/z value, which then strike the ion detector. The magnitude of the current
produced at the detector as a function of time (e.g., the physical field
in the mass analyzer is changed as a function of time or the time it takes
the ion to move a certain distance) is used to determine the m/z value
of the ion. This figure shows an ion trap mass spectrometer with ions
trapped in the electric field.

By changing the characteristics of the field, ions can be manipulated

and ejected from the trap to a detector. The time component
as a function of the field is what determines the m/z value of the ions.
Previous Page ||
Next Page
Back to Technologies ||
Back to Mass Spectrometry