DRX499 : VT SYSTEM UPDATE : PLEASE READ
Similar to the temperature limit test that was conducted on AV500, we completed a test to determine the maximum allowable sample temperature that is safe within the operational constraints of the DRX499 spectrometer. We have determined that +70 C will be the absolute maximum sample temperature that is sustainable without causing serious concerns, for an assumed data collection duration of 12 hours. If a rather long haul run is needed at 70 C beyond this duration, users should consult with the facility staff for viable methods.