UW CHEMISTRY NMR

Announcments & Equipment Logs

September, 2020

UPDATE ! AV500 – LOCK SYSTEM

We received some more input in the past few days about the Lock system being ‘unpredictable’.   But as it turns out, this was only an apparent observation, driven mainly by the miss-setting of the Lock field for different solvents.  The problem has been remedied and I have tested with two diverse lock solvents that are […]

Posted in Hardware, Spectrometer | No Comments »

UPDATE ! AV500 – LOCK SYSTEM IS WORKING FINE

Lock system is functioning as expected with a solvent such as CDCl3, which was reported to be a problem earlier today.    I have also checked with the aqueous solvent mixture of the decoy sample and locking is working OK now. NMR Facility Manager 9/20/2020

Posted in Hardware, Spectrometer | No Comments »

UPDATE ! AV301 : PROBEHEAD TESTED AND SYSTEM READY FOR USE

Dear Users The ATMA problem in 1H channel is fixed.  Given below is a 1D proton from a sample in CDCl3.  As is evident, the S/N as well as  lineshape are on the mark.   The only Caveat I will add is this : if you are trying to use an aqueous sample with 90% H2O, […]

Posted in Hardware, Spectrometer | No Comments »

UPDATE ! AV301 – PROBE IS REPAIRED AND WILL BE TESTED AND RELEASED SOON

The electronic issue has been rectified in-house.  I am hopeful that the probe should be available for regular use by tomorrow.   I will confirm the same in a following post before releasing the instrument. Thank you for your patience. NMR Facility Manager 9/3/20

Posted in Hardware, Spectrometer | No Comments »

UPDATE ! AV301 : PROGRESS BEING MADE IN FIXING THE PROBE

Hello Users Apropos the ailing probehead,  I am glad to add that the route cause is identified and we should be able to remedy it in-house. I hope my next update will be to say that the probe is back in action ! Stay Tuned NMR Facility Manager 9/3/2020

Posted in Hardware, Spectrometer | No Comments »

ALERT ! AV301 – SYSTEM DOWN DUE TO PROBEHEAD ISSUE

Hello All Many of you might have noticed and some of you have given feedback about the rather poor S/N on this system in the recent past. Especially, today people have noticed that ATMA is not working as expected. Further investigation shows that there is indeed a hardware issue with the probehead on the 1H […]

Posted in Hardware, Spectrometer | No Comments »