AV300, AV500: PRINTING OK
The long standing problem of not able to print spectra from either the Topspin window or with the xwinPlot, plot editor program is now resolved. For routine use, you can use Ctrl-P to print whatever is displayed in a active Topspin window to the printer, without the necessity of starting the plot editor.
Posted in Software, Spectrometer, Usage | No Comments »
DRX499 : VT SYSTEM UPDATE : PLEASE READ
Similar to the temperature limit test that was conducted on AV500, we completed a test to determine the maximum allowable sample temperature that is safe within the operational constraints of the DRX499 spectrometer. We have determined that +70 C will be the absolute maximum sample temperature that is sustainable without causing serious concerns, for an […]
Posted in Hardware, Spectrometer, Usage | No Comments »
AV300 : CLEANUP UPDATE
The AV300 system is functional. We are wrapping up running a few more tests. The spectrometer will be available for use from 11.00 am.
Posted in Hardware, Spectrometer | No Comments »
AV300 : NMR LAB FLOODING
An overnight flooding of the 1st floor of Bagley Hall, right above Rm.37, led to water leakage via the ceiling that drenched the AV300 console. The lab has been mopped up and we are checking that the console is not damaged due to this incident. The AV300 console will be released to regular use, latest […]
Posted in Hardware, Spectrometer | No Comments »
AV700 Probe Change
Please plan/prepare accordingly for the solid state MAS probe to be installed on September 17th, 2012.
Posted in Uncategorized | No Comments »