UW CHEMISTRY NMR

Announcments & Equipment Logs

ATTENTION ALL USERS ! : NMR LAB RM: 37 ACCESS RESTRICTION THIS WEEK

Facilities Works will redo the flooring of the corridor in Bagley Hall where the NMR lab is located  on the following days this week :  Tuesday, Wednesday and Thursday. This operation will last from 5.30 am until 8.30 am on each of these days.  During this time, no foot traffic will be allowed on the […]

Posted in Spectrometer, Uncategorized, Usage | No Comments

ATTENTION ! – USERS WHO EMPLOY J-YOUNG NMR TUBES…PLEASE READ !

J-Young NMR tube users must be fully aware by now that tubes that are bent from being perfectly linear cannot be ejected comfortably and they can even cause breakage inside the NMR probehead. Most of the time, the stuck tube entails complete removal of the probe from the magnet, followed by a not so simple […]

Posted in Hardware, Spectrometer, Usage | No Comments

AV700 : SOLUTIONS TO SOLIDS TRANSITION COMING UP – Dec 8

We will switch the AV700 to solids NMR beginning December 8th, Thursday and the system will serve the solids request queue until the end of December as per current estimates.  Based on how the runs pan out, we may be in a position to switch back to solutions even earlier.  Please look out for a […]

Posted in Spectrometer, Usage | No Comments

AV301 : TEMPORARY DATA DOWNLOAD DISRUPTION

AV301’s data tree won’t be available until 4 pm today, owing to computer maintenance.  Please note that NMR data download from all other hosts are not affected by this. Unless otherwise mentioned, the AV301 system will be back on line by 4 pm. NMR Facility Manager 11/22/16

Posted in Datasets, Host Computer, Operating System | No Comments

UPDATE ! DRX499 – SYSTEM READY FOR USE !!

The probe in the system now is : BBO- 1H{X}-Z gradient.   You can use the standard shim set : stdbbo before you shim up your specific sample. The probe that was damaged due to sample spillage will take a while to fix.  An update will be posted, once that probe is ready for test.  

Posted in Uncategorized | No Comments

UPDATE ! DRX499 – PROBE CHANGE IMMINENT

Since there was extensive sample spillage inside the TXO probe that is giving a large background interference, this probe cannot be employed in its current shape, presently.  We are  putting in the alternate probehead  i.e. BBO – X-{1H} Z Gradient and priming the same for routine use.  Current estimate is that we will release the […]

Posted in Hardware, Spectrometer | No Comments

ALERT ! DRX499 – INSTRUMENT DOWN DUE TO USER SAMPLE BREAKAGE

User sample tube has broken inside the probehead and the instrument is not usable currently. Probe needs to come out of the magnet to know the extent of damage or spillage. Based on the assessment, we will decide if the TXO probe can be back in the system or a different probe will take its […]

Posted in Hardware, Spectrometer | No Comments

UPDATE : AV300 – SYSTEM BACK IN USE – WITH THE OLD PROBE STILL

The ATMA Probe that was procured as part the hardware upgrade, unfortunately, had been found to have a faulty motor that prevents it from Automatic Tuning/Matching operation. The probe has been sent back for repairs and the older BBI probe has been put back and shimmed up. I will announce the completion of the upgrade […]

Posted in Hardware, Spectrometer | No Comments

AV300 – UPDATE : LOGIN PROBLEM RESOLVED.

No issues.  Please post a message, if you notice otherwise. NMR Facility Manager 10/24/16

Posted in Software, Spectrometer | No Comments

DRX499 – UPDATE ! SYSTEM IS BACK TO NORMAL

The hardware issue is fixed and Lock and Shim systems are tested to be ok. Please let us know if you observe otherwise. Thank you for your patience. NMR Facility Manager 10/18/16  

Posted in Hardware, Spectrometer | No Comments