Welcome to the Chemistry NMR Online Log Book – your  One Stop portal to communicate about the spectrometer to the Facility Staff in a timely manner. 

Please record your comments and relevant observations about AV500 instrument by submitting an entry using the “Post Log Entry” box found below.

Subscribe to AV500 logs

269 Responses to “AV500 Log Book”

  • richard says:

    some garbage

  • nicole rishwain says:

    Some peaks on the spectrum were upside down.

  • Kathleen says:

    Topshim failed repeatedly – field map error
    When tunning and matching, the match could never get low enough to meet the baseline – and was considerably high up
    Locked the sample repeatedly, lock line in the little window looked noisy

  • Victoria Zottarelli says:

    Had to rewobb 3 times and still did not work

  • Leo Zasada says:

    Spun rate read error prevented operation of software

  • James says:

    Will not shim with topshim (sometimes it works after the 5th attempt). Today it would not shim even after repeated topshim runs. (solvent was CDCl3).

  • Alex says:

    WOnt topshim

  • Delwin Elder says:

    I could not use the AV500. I got a pop up window saying “Spin rate read error” when I logged in. I could not eject the sample or enter any other commands

  • Leo Zasada says:

    Data comes out unusable. Topshim maybe working? Not any obvious errors but multiple fails between different samples. Running in DMSO (temp fine)