UW CHEMISTRY NMR

Announcments & Equipment Logs

September, 2022

AV500 : IMPORTANCE OF TOUCHING UP THE X,Y SHIMS

Dear Users Way back,  I had highlighted the importance of touching up the so-called transverse shims  a.k.a. X and Y shims, AFTER you have completed the Topshim command. I had shown a real-life example of the difference this simple step makes in getting a well resolved and great spectrum.   Here is a link to that post […]

Posted in Hardware, Spectrometer, Uncategorized, Usage | No Comments »

UPDATE ! LOGIN SYSTEM STABLE

Dear Users The root cause of login glitches noticed yesterday has been identified and a patch is applied.  If any of you still have login issues with your own Netid, please contact us by email immediately. Thank you NMR Facility Manager 9.23.22

Posted in Host Computer, Operating System | No Comments »

ALERT ! LOGIN SYSTEM UNSTABLE ON ALL NMR HOSTS

We are aware of the ongoing issue, where your UWNetid login allows you into some NMR instruments and not the others.  We are looking at this intensely and will try to fix this ASAP. Please wait for further updates. NMR Facility Manager 9/22/2022

Posted in Host Computer, Operating System | No Comments »

AV301 : SYSTEM READY FOR USE WITH SURROGATE BBI PROBE

Dear Users       The regular probe with a faulty ATMA motor assembly is pulled out for repairs.  Do not underestimate the probe that is placed in the magnet as a backup, because it is called Surrogate. This is a newly procured “gently used probe” in excellent condition and gives beautiful spectra as I show below. […]

Posted in Hardware, Probeheads, Spectrometer, Usage | No Comments »

IMPORTANT REMINDER ABOUT ATMA ! PLEASE READ

I am motivated to post this in the wake of ATMA problem with the AV301 probe but, this is generally applicable to both the 300 MHz systems and the new probes with ATMA capability we will get in the near future. Message :  Always LOCK first on your solvent BEFORE you attempt ATMA.  ATMA program […]

Posted in Hardware, Probeheads, Spectrometer | No Comments »

ALERT ! AV301 – PROBE DOWN DUE TO ATMA RELATED ISSUE

At this point, it is not clear if it’s a hardware problem or a  software related issue. We will try to resolve it as soon as we can.  If it comes down to the probehead hardware that might take more time to fix, I will place a backup probe in the system. I will post […]

Posted in Hardware, Probeheads, Spectrometer | No Comments »

GG500 : Addressing sporadic misbehavior of  ‘wobb’ Window

Users have noticed that the STOP button in the ‘wobb’ window gets grayed out sporadically.  When this happens, it appears that you cannot stop the process and proceed with the Acquisition.  This is not true.  I show below the pull down Sub Menu called User Specific Commands, which itself can be accessed from the main […]

Posted in Uncategorized | No Comments »

GG500 – BIT OF HISTORY, CURRENT ISSUES AND ROADMAP – PLEASE READ

GG500 – Apparent outlier in the pack : Before GG500, we had a really aging and tired DRX500 console, which needed to be driven with a software interface that was predictable and repetitive, to accommodate its fragile and irreparable hardware and produce useful NMR data. The curated work flow of IconNMR’s Routine Spectroscopy interface was […]

Posted in NMR Facility Usage | No Comments »