UPDATE ! AV300 – PENDING ISSUES RESOLVED – SEE DETAILS
First of, the system is ready for use ! NETWORKING ISSUE : The temporary network glitch is resolved now and users can login with their UWNetid and store data in the UDRIVE. If you notice anything different with your account, please send me an email. SAMPLE LANDING AND SPINNING OK : Lot of debris and […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV301 – SAMPLE HANDLING AND LOCK WORKING OK
The issue of BSMS not responding, leading to sample handling and lock was a temporary issue, probably caused by a voltage surge. The issue is resolved and the system is functioning normally now. NMR Facility Manager 2/12/2019
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV500 – THE TBI PROBE IS BACK
The mechanical issue compromising the 1H match has been fixed in-house and the probehead is back in the magnet. We have replaced the broken 1H match knob with a homemade knob. I will show a picture below that clearly identifies the new homemade knob. Only for 1H Match, you need to use the following tune/match tool […]
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV300 – SAMPLE LANDING ISSUE – UPDATE
Sample spinning with this instrument is not working and this is also related to the BSMS system’s observation that sample is ‘missing‘. We are looking into most plausible causes before we decide to pull the probe out, to minimize the down time with the instrument. Please note that the only alternate probehead that can […]