AV800 Log Book

Welcome to the Chemistry NMR Online Log Book – your  One Stop portal to communicate about the spectrometer to the Facility Staff in a timely manner. 

Please record your comments and relevant observations about AV800 instrument by submitting an entry using the “Post Log Entry” box found below.

Subscribe to AV800 logs

5 Responses to “AV800 Log Book”

  • Jordan Anderson says:

    Topspin refuses to open, and responds with the error: “java.io.IOException: Disk quota exceeded”

  • Jordan Anderson says:

    We are unable to find our data on the server. The AV800 folder on the phoenix.chem.washington.edu site appears to be empty.

  • Liladhar Paudel says:

    Sample stuck at the top of the magnet leg. The new sample inserted through the auto-sampler was taken to the top and stuck there (stuck at the intersection between horizontal and vertical path to the magnet). So I need assistance to resolve this problem as soon as possible. Thanks.

    –Liladhar Paudel

  • Liladhar Paudel says:

    While running gHSQC I found that all of sudden gain changed from 2050 to 1 when I did rga. Then I tried to topshim (because I thought shimset might have lost). when try to shim using topshim, it was doing something at the beginning but finally stops with an error message ” RawCheck – the Z gradient does not work” in a popup window.

  • Jordan Anderson says:

    There was some problem with the sample lift. When I ejected the previous sample it crashed into the bottom of the sample holder and broke. There doesn’t appear to be any lift gas flow.