UPDATE ! AV500 – GOOD OLD BBI PROBE IS BACK…
We have managed to fix the X-channel in house and also replaced broken parts where the sample resides inside the probehead. Since things look pretty good with the probe (see image of water suppression : BBI_Sucrose_ES_after_fix ) we are releasing the instrument for use with BBI probe in place. We will check a few more things with the probe but my investigation thus far shows that the probe is ready for routine spectroscopy.
I must add this point the nature of the damage sustained on the X-channel. The probe suffered, what is known as ‘arcing’, which pretty much burnt out one of the capacitors. There may be more than one reason for such a damage to occur. But, a user, carelessly setting up an experiment where a large amount of pulse power-pulse duration combination or high decoupling power is employed, can certainly lead to such a damage. If you ever have any doubt about the experiment, please contact me or Adrienne for assistance. Don’t second guess your parameters.
Thank you for your cooperation.
NMR Facility Manager
5/12/16