UW CHEMISTRY NMR

Announcments & Equipment Logs

UPDATE ! AV301 – X -CHANNEL WORKING AS EXPECTED.

We have tested both 13C and 31P observe with 1H decoupling and the transient spikes that were observed earlier is not present anymore.   The hardware is working as expected and a mis-adjustment in the parameter setup is identified as the main cause of the noise.

We shall keep a close watch, nevertheless.

Please note that you do not need anything more than 300 ms of AQ when you run 13C or 31P with decoupling.  The inherently large spectral resolution will afford a well resolved spectrum even with this FID length.

On the other hand, increasing the acquisition time beyond this with broadband decoupling ON, will lead to probe damage.  Please consult one of us, when in doubt. If you are convinced that a longer FID length is needed, there are workarounds to achieve that.

Thank you for helping to preserve your spectrometer.

NMR Facility Manager

2/27/2018


Leave a Reply

Your email address will not be published. Required fields are marked *

The maximum upload file size: 3 MB. You can upload: image. Links to YouTube, Facebook, Twitter and other services inserted in the comment text will be automatically embedded. Drop file here

Item Categories

Archives