UW CHEMISTRY NMR
Announcments & Equipment Logs
REMINDER ! OBSERVING X-NUCLEI IN AV500 and DRX499
Please remember the following. You should make sure that the X-channel cable is connected to the correct nucleus port on the probe head, before you tune/match or observe NMR signal from that nucleus.
For instance, on AV500 :
- If you want to observe 13C, make sure that the RF cable is connected to the 13C PORT (BLUE COLOR) on the probehead.
- If you want to observe 31P, make sure that the RF cable is connected to the 31P PORT (RED COLOR) on the probehead.
Similarly, on the DRX499, you should make sure that the RF cable is connected either to 13C or 19F to run those respective experiments.
As always, feel free to ask, if you have questions.
NMR Facility Manager
3/9/2018
Leave a Reply