UW CHEMISTRY NMR

Announcments & Equipment Logs

UPDATE : AV700 – VT CALIBRATION COMPLETE

With the backup chiller in place, the VT calibration for the high temperature range is complete.  Now, you can simply request the temperature you need to see at the sample.

The system will automatically adjust the temperature measured at the sensor to achieve this.

I show below an example to demonstrate the accuracy achieved.  The requested temperature is 310 K and the actual temperature, as measured by the Delta Chemical Shift of the High VT standard 80% Ethylene Glycol in DMSO-d6 is 310.09 K, pretty close to the demanded value.

NMR Facility Manager

5/15/2019


Leave a Reply

Your email address will not be published. Required fields are marked *

The maximum upload file size: 3 MB. You can upload: image. Links to YouTube, Facebook, Twitter and other services inserted in the comment text will be automatically embedded. Drop file here

Item Categories

Archives