ALERT ! AV500 OK – WITH IMPORTANT CHANGE – PLEASE READ !
CURRENT STATUS :
First I will highlight that the sporadic Topshim problem is gone now and the shimming is behaving as expected. That should come as a relief to all of you who exclusively use 2H lock solvents other than (H2O+D2O) or (CH3OH+D2O).
IMPORTANT – PROBEHEAD CHANGED TO BBI
In the process of system troubleshoot, I swapped in the BBI probe and pulled out the existing TBI probe. This probe is also optimized for 1H sensitivity and gives excellent signal on the samples I ran. You can run Topshim on this as before. And, as before, you have to tune/match this probe by hand – NO ATMA !
CAPABILITIES PRESERVED :
You are still getting a 1H sensitive broad band probe that should handle all your X-nuclei as well as 1H needs. The probe is also in good condition for all your VT experiments.
The only change you will notice is that you have to tune/match the 13C channel as in the old days i.e. using the Tune/Match card and dial in 4 digits for the 13C tuning and 3 digits for the matching. Same is true for all the X-nuclei. If you have doubts please contact Adrienne or myself to give you a hand.
Further Thoughts :
The TBI probe itself deserved some much needed attention on the bench, outside the magnet for a while now. The 13C tuning/matching has been flaky and occasionally people noticed some 1H channel anomalies. There are a few mechanical spruce ups that are in order.
Using this transition, we will work to give a fresh look/feel to the TBI probe and return it to service in the next month or so.
THE INSTRUMENT IS RELEASED FOR IMMEDIATE USE !
NMR Facility Manager
6/18/2021