UPDATE ! NEO500 – PROBE CHANGED TO TXO-19F{1H,13C} – PLEASE NOTE
Dear Users
The previous TBI probe had to be removed for a service operation, and I have placed the TXO-19F PROBE for the present as an alternative. Please note the following :
- Nuclei : 19F, 1H and 13C
- You can run 1H decoupled 19F experiments
- To switch between 19F and 13C, a cable swap is needed (please ask for my help)
- By default, the system will be connected to run 1H and 13C.
- Feel free to contact me if you want to run 19F and I will show you how – its a simple process.
- The pulse widths for the 1H and 19F are optimized.
- TODO :
- I am yet to complete the optimization of the shim matrix, and so Topshim will throw out errors. Workaround: Please apply the same simple low order hand shimming process that you use in 300, here also. Along with spinning of the sample, you will get a decent 1D spectrum.
- I released the instrument to meet demand, but the sundry optimization steps will be completed by Monday in intermittent blocks of time.
- Optimization of 13C p90 pulse is yet to be completed, but the default value provided will give you reasonable 13C spectrum.
- Additional shim files, specific to different lock solvents, will be created on Monday.
Thank you for your patience and cooperation.
NMR Facility Manager
7.25.2025