UW CHEMISTRY NMR

Announcments & Equipment Logs

UPDATE – NEO500 – PROBE IS FULLY FUNCTIONAL WITH TOPSHIM

Dear Users

The probe is optimized for different lock solvents that should make shimming with Topshim quick and easy.

  • I have created individual shim files that are the optimal starting points for the most often used lock solvents.
  • The files are named as below :
    • txo.water (for aqueous samples, including those in D2O)
    • txo.acteone
    • txo.cdcl3
    • txo.c6d6
    • txo.dmso
  • After you lock on your solvent, please enter : rsh txo.<your solvent> 
  • Type topshim tunea
  • Please enter ‘wobb’ and manually tune and match the probe.
  • Run rga prior to acquiring data using zg.

I will add more solvent dependent shim files to the above list, in due course.  If you urgently need a solvent not in the list, that needs to be added to the line-up, please contact me.

NMR Facility Manager

7.28.25