UW CHEMISTRY NMR

Announcments & Equipment Logs

AV301 Log Book

Welcome to the Chemistry NMR Online Log Book – your  One Stop portal to communicate about the spectrometer to the Facility Staff in a timely manner. 

Please record your comments and relevant observations about AV301 instrument by submitting an entry using the “Post Log Entry” box found below.

Subscribe to AV301 logs

257 Responses to “AV301 Log Book”

  • Adelle says:

    When trying to eject dummy sample I am getting the error “spin rate read error” repeatedly

  • Devin Rollins says:

    When I tried to run “topshim” I was prompted with a message saying “Please state, how the shim system is attached to the shim current boards…” and was required to choose one of two options in order to continue.

  • Kiyoshi Colon says:

    Topshim command failed due to apparent low signal-to-noise ratio.

  • Jay Lee says:

    ATMA doesn’t work. The “Dip” in the Wobble curve moves too much for the instrument to tune/match properly.

  • Rajan Paranji says:

    Log Test Comment – ignore

    When it shines it pours…

  • Connor says:

    Cannot eject dummy sample – error “spin rate read error”

  • Matt says:

    Cannot eject dummy sample – error “spin rate read error”

1 24 25 26


Leave a Reply

Your email address will not be published. Required fields are marked *

The maximum upload file size: 3 MB. You can upload: image. Links to YouTube, Facebook, Twitter and other services inserted in the comment text will be automatically embedded. Drop file here

Item Categories

Archives