UPDATE ! AV301 : PROBE SWITCHED TO TXD-1H (13C,19F) – NO GRADIENTS
Dear Users
As I announced last week, the ailing ATMA probe has been pulled out and in its place, the good old TXD probe is placed and primed up. This probe has three dedicated channels for 1H, 13C and 19F.
THIS IS NOT A BROADBAND PROBE !
This means that you cannot observe anything other than the above three nuclei.
THERE IS NO Z-GRADIENT !
The consequence is that you must :
- load the standard shim file first : rsh stdtxd
- Spin the sample
- Touch up at least the Z and Z2 shims
We are putting all possible efforts to fix the PABBI probehead that can do Automatic Tuning and Matching as well as Gradient shimming.
As soon as that probe is ready for use, it will get back in action, rest assured.
NMR Facility Manager