Welcome to the Chemistry NMR Online Log Book – your  One Stop portal to communicate about the spectrometer to the Facility Staff in a timely manner. 

Please record your comments and relevant observations about AV301 instrument by submitting an entry using the “Post Log Entry” box found below.

Subscribe to AV301 logs

257 Responses to “AV301 Log Book”

  • Ruining Yang says:

    Unable to lock on CDCl3

  • Brian Shn says:

    Would not lock on CDCl3

  • Austin shaff says:

    Would not lock CDCl3

  • Bradley Gardner says:

    Would not lock on C6D6

  • Micaela Homer says:

    Would not topshim on a sample in DMF-d7, said signal to noise too low. Also, the first time I tried ATMA, got the error :channel can’t be wobbled with this probe

    I typed ATMA again and it worked the second time.

  • Sarah Zeitler says:

    It’s taking multiple attempts to lock cdcl3. I’ve been tracking with lockdisp to make sure it locks before I shim and it just says job finished without actually locking to the solvent.

  • Jack says:

    Cannot topshim in CD3CN. Repeating attempts using the same sample gave different errors: signal-to-noise, too many points lost, BSMS issue.

  • Kathleen says:

    Lock said it completed successfully, but topshim said that lock was needed. Redoing the lock fixed this

  • Morgan Skala says:

    Was able to shim but shimming didn’t work well at all.

  • Kathleen says:

    topshim failed repeatedly w/ CD3CN solvent – “field map – signal-to-noise is too low