Welcome to the Chemistry NMR Online Log Book – your  One Stop portal to communicate about the spectrometer to the Facility Staff in a timely manner. 

Please record your comments and relevant observations about AV301 instrument by submitting an entry using the “Post Log Entry” box found below.

Subscribe to AV301 logs

261 Responses to “AV301 Log Book”

  • Kathleen says:

    Constant spin rate read errors made it almost impossible to inject/eject samples.

  • Ella Spurlock says:

    When I was trying to take 1H NMR at 11:40 AM, topshim kept failing. The error is “FieldMap – signal to noise is too low.” This happened even after I ejected, re-locked, and tried again. And also after I tried “rsh stdpabbi”.

  • Brad Cleveland says:

    Topshim command failed due to apparent low signal-to-noise ratio.

  • Adelle says:

    When trying to eject dummy sample I am getting the error “spin rate read error” repeatedly

  • Devin Rollins says:

    When I tried to run “topshim” I was prompted with a message saying “Please state, how the shim system is attached to the shim current boards…” and was required to choose one of two options in order to continue.

  • Kiyoshi Colon says:

    Topshim command failed due to apparent low signal-to-noise ratio.

  • Jay Lee says:

    ATMA doesn’t work. The “Dip” in the Wobble curve moves too much for the instrument to tune/match properly.

  • Rajan Paranji says:

    Log Test Comment – ignore

    When it shines it pours…

  • Connor says:

    Cannot eject dummy sample – error “spin rate read error”

  • Matt says:

    Cannot eject dummy sample – error “spin rate read error”