  • Brandon Kier says:

    multi_zgvt, multizg, and related macros non-functional. Cannot reset temperature at end of run, or run multiple experiments in sequence. Is an alternative command functional?

  • Carolyn Valdez says:

    On proton channel, the tune and match is shifting significantly, even over short periods of time. When I re-checked wobb after 10 minutes, I couldn’t even see a parabola anymore…

  • Brandon Kier says:

    The instrument is not cooling below ~293 K. Low-temperature runs are not possible.

  • Rajan Paranji says:

    Here is another test message for Adrienne. Hi ! This is Thursday the 20th June, a cloudy day.

  • Kathleen Olivas says:

    printer is still not connected to the instrument.

  • Kathleen Olivas says:

    printer is not connected or set as the default to actually be able to print anything… I thought the printer was supposed to be working now?

  • Paranji says:

    Sam
    H2O+D2O will give a FWHM anywhere between 10 to 20 Hz. This sample is not a good one to test lineshape issues. One of us will take a look.

  • Samantha Connelly says:

    The shims are fairly atrocious this afternoon. After 20 minutes of shimming (starting from the standard shim file), my line widths still aren’t better than 6-7 Hz (full width, half-height). The sample is homogeneous (and gave a much better spectrum this morning on the 301 and this instrument).

    To double check that my sample wasn’t the issue, I’ve just loaded the standard shim file and recorded a spectrum of the dummy sample (H2O/D2O). This too has line width around 8 Hz (fwhh). This seems fairly poor?

  • Adrienne Roehrich says:

    The lock/lockdisp issue has been resolved.

  • Sophia Tran says:

    I received an error message after typing “lockdisp” that read “Error opening spect:/dev/tty03 locked by process 8655” After pushing close, another error came up reading “No acquisition hardware available for:: 128.95.172.145”