UW CHEMISTRY NMR
Announcments & Equipment Logs
September, 2017
ALERT ! AV500 – SYSTEM NOT FULLY FUNCTIONAL
An issue with the 3rd RF channel is being looked into. Without this channel, Topshim will not work. I am hoping that its a quick fix. If not, I will give you an update. Stay tuned. NMR Facility Manager 9/18/17
Posted in Hardware, Spectrometer | No Comments »
ALERT ! ALL NMR SYSTEMS RESET AFTER POSSIBLE POWER DISRUPTION
Though I haven’t heard of specific reports, all the NMR consoles in Rm:37 evidently underwent some voltage fluctuation, possibly anytime over the weekend. All the log posts of today morning are related to this. The systems have been reset to their usable condition now. NMR Facility Manager 9/18/2017
Posted in Hardware, Spectrometer | No Comments »
UPDATE – DRX499 – LOCK SYSTEM IS RESET
I have checked that the locking on CDCl3 as well as other solvents is functioning as expected. Nevertheless, do post a message in the log if you notice otherwise, especially with non-standard solvents. NMR Facility Manager 9/6/17
Posted in Hardware, Spectrometer | No Comments »
ALERT ! SYSTEMS RESTORED – AFTER POWER FLUCTUATION
The apparent Lock anomaly noted with AV300 has also been found with AV301 and possibly DRX499. Apparent line voltage fluctuations were reported today and similar to last week, that could be cause of the Lock disruption. The 300-s are restored to normal operation as we speak and I am checking that DRX499 is functioning as […]
Posted in Hardware, Spectrometer | No Comments »