UW CHEMISTRY NMR

Announcments & Equipment Logs

IMPORTANT UPDATE ! AV300 – THE ATMA PROBE IS BACK IN ACTION !

Dear Users

Your patience has paid off and at last we got the probe fixed (again) and this time things seemed to have fallen in place.   We are so confident of this that, without any further delay, we release the instrument with the PABBI ATMA probe for your immediate comfort and producitivity.

So, remember :

  • Lock the probe with the correct solvent name first.
  • type ‘atma’  to tune/match the probehead for the solvent on which you locked.
  • Start with ‘rsh stdpabbi’  to be 100% sure that you are beginning from a decent shim set. This makes Topshim converge faster
  • Start Topshim.  Enjoy !

Please let us know via the logbook page if you notice anything that needs attention :

NMR Logbook – av300

NMR Facility Manager

2/22/2021