UPDATE ! AV500 – INSTRUMENT USABLE WITH LIMITED FUNCTIONALITY
We have ordered the necessary replacement parts and asked the manufacturer to ship them asap. I am hoping that it will be Thursday before we get these parts. I am releasing the instrument for nominal use, with the following caveats, until we get the new boards in :
- The wobb signals WILL be noisy, but the probe can be tuned ok.
- The pulse width for 1H is really large i.e. 18 us at -2 dB power. Contrast this with something like 8 us at lower power.
- The large pulse width also means that 1H decoupling may not be very efficient.
- Those who attempt spin locking experiment such as TOCSY SHOULD BE EXTRA CAREFUL not to over drive the probe. PLEASE ASK IF YOU ARE IN DOUBT !
The new boards should remedy all these and I am optimistic that by Friday the system should be in prime condition and ready for use.
NMR Facility Manager