Was eventually able to run a proton experiment, but took approx. 15 minutes due to slow computer issues. Almost unusable.
Computer running slow…omputer took 7 mins to load the proton experiment…ran out of time to run sample.
The computer was very slow today. Program took more than 10 minutes to load and a script error popped up. Unable to run sample.
As of today i.e. Monday the 18th, the said problem is not reproducible. The system seems to be functioning normally, after the disk clean up last week. We will keep a close watch.
Computer running extremely slow (over 15 min for the program to launch)…unable to run sample.
Home directory is getting full, especially with user bookmarks and temporary files via the web browser. I am working on a cleanup operation right now. I will increase the disk capacity in the coming days.
Topspin slow and log book slow to load after logging in. Getting a “scripts stopped running” error or something like that. Topspin also slow to react to commands, clicks, etc.
We will replace the broken depth gauge today. Please borrow the one from AV301, in the meantime.
Problem with phoenix is already fixed..
the NMR tube depth gauge broke
unable to access data on the phoenix server for the AV300