AV301 : SAMPLE SPINNING ISSUE

A possible blockage in the pneumatic system is causing the sample not to spin correctly.  Related to this is the fact that, the BSMS keyboard might indicate the sample as ‘missing’, even when the sample is seated inside.    The shim stack needs to be removed to clear this problem. Users are requested not to use […]

Posted in Hardware | No Comments »

AV500 : automation scripts not working

Familiar automation scripts such as ‘xaua’ are not functioning as expected.  This is being looked into.  Please use workarounds that are viable. For instance, ‘xaua’ can be replaced with ‘rga’ followed by ‘zg’.   Please contact facility staff for assistance with this.

Posted in NMR Applications, Software, Usage | No Comments »