IMPORTANT REMINDER ABOUT ATMA ! PLEASE READ
I am motivated to post this in the wake of ATMA problem with the AV301 probe but, this is generally applicable to both the 300 MHz systems and the new probes with ATMA capability we will get in the near future. Message : Always LOCK first on your solvent BEFORE you attempt ATMA. ATMA program […]
Posted in Hardware, Probeheads, Spectrometer | No Comments »
ALERT ! AV301 – PROBE DOWN DUE TO ATMA RELATED ISSUE
At this point, it is not clear if it’s a hardware problem or a software related issue. We will try to resolve it as soon as we can. If it comes down to the probehead hardware that might take more time to fix, I will place a backup probe in the system. I will post […]
Posted in Hardware, Probeheads, Spectrometer | No Comments »
UPDATE ! GG500 – INSTRUMENT BACK TO NORMAL
Dear Users The sample handling issue that was reported had since been resolved. The instrument is open for reservation but for a cryogen fill slot this morning. Thank you for your kind understanding. NMR Facility Manager 07/08/2022
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV500 – SYSTEM SWITCHED BACK TO 19F CAPABILITY
Hello Users The Probe now In : TXO-19F,13C {1H} – Z gradient Please load the Standard Shim set by doing this on the Topspin command line : rsh stdtxo (In case you are curious, the nucleus/nuclei inside the curly braces is/are the so-called ‘decouple’ nucleus. The probe is typically designed to give the best sensitivity […]
Posted in Hardware, Spectrometer | No Comments »
AV500 : PLEASE NOTE THE TEMPORARY PROBE SWAP
Dear Users We swapped the probe today to BBO as a one time process to access a rare nucleus. Those of you who will use the instrument over the weekend, please note that, this a BBO Probe that includes 1H and X-nuclei coverage, that includes 13C. What you WON’T have is 19F capability. We will […]
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE ! 500 NMR-s WORKING OK
Dear Users Most probably, there was a power fluctuation in the past day or so that led the 500 NMRs into an unusable state. The situation is remedied and everything looks normal with these instruments. In case you are curious, the 300-s have been behaving OK when I checked this morning. NMR Facility Manager 5/31/2022
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV300 – ATMA NOISE ISSUE IS ADDRESSED
Dear Users Some of you had notified us that the V-shaped ‘wobb’ curve was showing lot of noise and ATMA was having issues in tuning/matching automatically. The problem was identified to be localized only to the 1H channel and subsequently, the cause for this has been addressed. My tests today show that the system is […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE : AV500 – RELATED
Dear Users We have been receiving sporadic log posts about the Spin Rate Error and occasional Locking related issues. As I write this, the system is in a normal state and I have recalibrated the sample spinning system and tested it. Locking per se works fine but I have to see if specific solvents are […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE – AV500 ! TXO-1H {13C,19F} PROBE IS BACK
Dear Users As we suspected, a mechanical issue was the cause of 13C channel’s failure. We could fix this in-house with the happy ending that 13C Tune and Match behave pretty smooth now. The TXO probe is back in the magnet, ready to collect data. NMR Facility Manager 12/7/21
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV500 – POSSIBLE PROBE SWAP TODAY ON THE CARDS
Dear Users The TXO probe has been needing some attention vis-a-vis the 13C channel. As I speak, we are looking into it. If it’s a quick fix in-house, the same probe will go back in the instrument. If not, I will have to place the BBO probe in its place. Sorry for this last minute […]
Posted in Hardware, Spectrometer | No Comments »