DRX499 : UPDATE – OPERATIONAL WITH LIMITED FUNCTIONALITY
The spectrometer with the TXO probe is operational fully with 1H and 13C channels. The 19F is not yet ready for prime use as I write this and I will post an update once that issue is resolved as well. The instrument is open for regular use, with the above caveat.
Posted in Hardware, Spectrometer, Usage | No Comments »
ALERT ! : PROBE CHANGE – AV500 and DRX499
Probe Swap date : August 1st, Friday AV500 : The existing TXO probe will be replaced by the Z-gradient BBI probe and primed for use. The system should be ready for use before the day is over. DRX499 : The TXO-19F probe will take the place of the current probe. Due to unavailability of gradient […]
Posted in Hardware, Spectrometer | 1 Comment »
UPDATE ! DRX499 – SYSTEM BACK TO NORMAL
Please let us know if you face any issues again.
Posted in Hardware, Spectrometer | No Comments »
UPDATE – AV700 : ISSUES RESOLVED
Acquisition as well as Lock problems reported earlier have been fixed. Please send us a feedback if you still notice any issues with as much information that you can provide.
Posted in Hardware, Spectrometer | No Comments »
UPDATE : DRX499 – AVAILABLE FOR USE
Many of the issues on the drx499 users have noticed in the past two weeks were due to a faulty board. This board has been replaced and the console is now fully functional. A test run shows normal operation. Please contact Adrienne (amr0311@uw.edu) or Rajan (paranji@chem.washington.edu) with questions.
Tags: drx499
Posted in Hardware, Spectrometer | No Comments »
UPDATE : DRX499 – AWAITING REPLACEMENT CIRCUIT BOARD
A faulty ADC board is recognized to be the most probable cause of problems and an order for its replacement was placed with instructions for expedited shipping. We will update the status once the board arrives and we test it out in the system.
Posted in Hardware, Spectrometer | No Comments »
ALERT ! : DRX499 – SYSTEM DOWN – HARDWARE MALFUNCTION
The receiver amplifier is suspect and is giving rise to glitches along the spectral range. Some users had already reported the ‘wobb’ curve showing peculiar noisy features that is also traced to this cause. The replacement/repair options of the board are being explored and we will try to fix this as soon as feasible. Until […]
Posted in Hardware, Spectrometer | No Comments »
AV301 : NOTE On PROBE CHANGE
Following the stuck sample tube incident, a troubleshooting attempt of the ATMA – PABBI probe led to an unfortunate damage to the probe and it is being sent to Bruker for repairs. This is the reason behind the hurried replacement in this instrument, with the TXO-19F {1H,13C} triple resonance probe. If you have further questions […]
Posted in Hardware, Spectrometer | No Comments »
AV500 – TOPSHIM – NOTE ON AUTOMATIC SHIMMING
There are still some pending calibration operations with the TXO-19F probe that was put in yesterday and this makes the Topshim automatic shimming unusable currently. As you may be aware, there’s been an impact on available NMR time due to probe issues on other instruments. To minimize this, we will keep using the TXO-19F {1H,13C} […]
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE ! DRX499 – LOCK SYSTEM RESTORED
Users should be able to lock on the solvents as usual. Please use the default shim file stdbbi and shim up at least the minimal lower order shims to get a good lineshape.
Posted in Hardware, Spectrometer, Usage | No Comments »