AV500 : IMMINENT PROBE CHANGE

The current TXO-1H,19F and 13C  probe has been identified with a inter-channel cross talk problem.  The probe is functional in and around room temperature,  but for those users doing  low temperatures, they will find it difficult to ‘wobb’ the probe at these temperatures. The TXO probe will be replaced in approximately a week’s time with […]

Posted in Hardware, Spectrometer | No Comments »

AV500 : LOCK UPDATE

Users had noted that the intermittent lock issue was back again.  A careful scrutiny was made and a needed configuration change was made to solve this problem.  The system should lock on all solvents without any error now.  If it is otherwise, please do note the same in the Log Book along with any other […]

Posted in Hardware, Software, Spectrometer | No Comments »

AV500 UPDATE : LOCK ISSUE FIXED

A possible power fluctuation over the weekend led to the disrupted lock and acquisition conditions of the instrument. These are now resolved and users can continue with normal operation of the instrument.

Posted in Hardware, Spectrometer | No Comments »

ALERT : PRINTER IS FUNCTIONAL AGAIN

Users can either print directly from Topspin or use the ‘plot’ program to direct print requests to the Laserjet printer, as before. Please keep facility staff informed if you notice any anomaly.

Posted in Hardware, Spectrometer | No Comments »

ALERT : NMR PRINTER OUT OF ORDER

An irrecoverable hardware problem has been found with NMR room printer.  Until we procure an alternate solution, users are requested to do their printing in their lab/office printers.  You can create printable  .jpg or .pdf images of the spectra from the NMRJuicer computer and send the same to your email account OR store in a […]

Posted in Hardware, Spectrometer, Usage | No Comments »

AV301 UPDATE : 1H {13C} PROBE WITHOUT VT VARIATION

This old double resonance probe cannot do VT variation.  The lineshape as it exists is usable for the routine runs but needs further optimization.  Since this probe will be treated only as a temporary resource, we are releasing the instrument as it is, to reduce the deficit of dysfunctional instruments to one i.e. AV500.   PLEASE […]

Posted in Hardware, Spectrometer, Usage | No Comments »

REMINDER : NMR FACILITY DOWN DURING PLANNED POWER OUTAGE

There will be a planned power outage in Bagley Hall this Saturday, the 18th, which will affect the NMR facility instruments as well.  The instruments will therefore be turned off by 4.00 pm, Friday the 17th and restored after the electrical maintenance is completed.

Posted in Hardware, Spectrometer | No Comments »

AV500 : UPDATE – DAMAGED HARDWARE

An issue has been identified with the RT shim stack and this makes the instrument unusable currently.  Since repair of this part is not feasible in house, we have to wait for a replacement of the same.   Users will be updated on that front when more information is available.

Posted in Hardware, Spectrometer | No Comments »

ALERT ! AV500 DOWN…

Hardware malfunction is being investigated.  The instrument will be down possibly until tomorrow.  Further updates will be posted here.    

Posted in Hardware, Spectrometer | No Comments »

AV800 : UPDATE – SYSTEM STILL UNSTABLE

Thorough investigations thus far indicate two possible points of instability in the system that leads to sporadic and unpredictable failure of the spectrometer. Since a troubleshoot of this nature entails long periods of test runs and observations, the system is not ready to collect useful data yet.   We will keep users posted as to any […]

Posted in Hardware, Software, Spectrometer | No Comments »