UW CHEMISTRY NMR
Announcments & Equipment Logs
Spectrometer
ALERT ! : NMR RESERVATION SYSTEM MALFUNCTION
We are aware that you are not able to reserve NMR time via the usual webpage front end. We are looking into the reservation server as we speak. I will post an update as soon as the problem is resolved. If it turns out that the problem will take longer than 30 minutes resolve, we […]
Posted in Spectrometer, Usage | No Comments »
ALERT ! AV500 – PROBE SWAP – CHANGE IN SCHEDULE
The original schedule of Probe swap in this instrument has been advanced to today/tomorrow. Part of the motivation for this is the (good) news that the REVIVED DRX499 is almost ready for full scale deployment within a few days. We are placing a new probehead called TBI-1H,13C {X} – Z gradient in AV500. This probe […]
Posted in Hardware, Spectrometer | No Comments »
IMPORTANT NOTE : UPCOMING TRAINING ON MANUAL SHIMMING AND TUNE/MATCH
Currently, we have only two spectrometers equipped with the ATMA facility and Topshim functionality. Decidedly, both these features take the drudgery out of routine NMR run, especially short ones. If you are someone who come to totally rely on these features for your NMR runs, you do have to face the following cons : You […]
Posted in Spectrometer, Usage | No Comments »
ALERT ! AV500 : IMMINENT PROBE CHANGE NEXT WEEK
The BBO probe will be installed and tested and hopefully be ready by the forenoon of 29th Nov ’18, Thursday. Please take note of this and plan accordingly. NMR Facility Manager 11/29/2018
Posted in Hardware, Spectrometer | No Comments »
AV700 and DRX499 UPDATE : PLEASE READ – SOLUTIONS TO SOLIDS TRANSITION – PLANS
The AV700 NMR spectrometer will be receiving a new Solid State MAS NMR probehead within the next few weeks. I will announce the finer details of this piece of equipment later. Please note that, currently the instrument is transitioned to solid state NMR mode to confirm its functionality before the new probehead is installed and […]
Posted in Hardware, Spectrometer, Usage | No Comments »
ALERT ! AV500 – TOPSHIM GLITCH
Topshim may tend to fail for low sensitivity lock solvent such as CDCl3, but it will work fine with solvents such as C6D6, CD3CN or MeOD-d4. We will be looking to resolve this by engaging the instrument in an intermittent fashion to minimize down time. An update will be posted once the issue is fully […]
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV500 – PROBE CHANGED TO TXO- 1H {19F, 13C}
Since the existing BBI probe had suffered a breakage in the VT dewar assembly, we were compelled to abruptly change the probehead to TXO- 19F probe. There was an original plan to do this probe swap by next week, in any case. Once the probe vitals are checked it will be released for regular use. […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV300 : SYSTEM BACK TO NORMAL
1H ACQUISITION Works as expected. Pulse width is already optimized. If you are in doubt you should run command “getprosol”. X ACQUISITION (13C, etc.) Works as expected. You might notice an increased noise level if you happen to run the spectrum on a previous sample for which you have data and compare it. The 90 […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV300 : ONGOING HARDWARE INVESTIGATION
Backdrop : We found a faulty RF circuit board serving the 1st channel of this instrument about a week ago but it had not completely stopped working. A swap of this unit with an identical 2nd unit was done to keep the instrument working at least for 1H observe. Current status : Notwithstanding this, the […]
Posted in Hardware, Spectrometer, Uncategorized | No Comments »
AV500 NOTE : TOPSHIM IS WORKING AGAIN
Based on several user samples tested as well as a few standards I tried myself, the Topshim program is working as expected. Please go ahead and try it yourself and give a feedback if it fails on your sample. If a failure occurs, please provide as many details as possible, such as the solvent used, […]
Posted in Hardware, Spectrometer | No Comments »