DRX499 – UPDATE : SYSTEM USABLE AGAIN

Apologies for the delay but the hung process has been cleared and users can start logging in to the instrument as before. NMR Facility Manager 10/15/15

Posted in Spectrometer, Usage | No Comments »

UPDATE : AV301 : PROBE DAMAGED

The 1H tuning capacitor seems to have been forced beyond its limits and the probehead is unusable currently.  We are working on a replacement probehead to take its place. The system will not be ready until tomorrow noon.  Please watch this space for further updates.  

Posted in Hardware | No Comments »

ALERT: av301 TUNE & MATCH ISSUE

We are currently looking into an issue with the 1H tune and match on the av301. We will update when the issue is resolved.

Posted in Hardware, Spectrometer | No Comments »

ALERT : DRX499 – PROBE CHANGE EVENT – 16th Oct ’15

The currently installed TXO 19F {1H,13C}  probehead will be replaced by the BBO – X{1H} Z gradient broad band probe beginning 16th October, Friday.  The DRX499 needs considerably more time to prime the shims and so the system will be ready for regular operation on Monday, the 19th.   We will keep you informed of […]

Posted in Hardware, Spectrometer | No Comments »

AV301 : FOLLOWING UP….

I want to highlight the optimized state of the TXD probe by showing you a representative 1H spectrum view below.  This should encourage users to feel confident about the results of our optimization efforts and hopefully users can reap the full benefits that AV301 has to offer. The full 1H spectrum shown here is from […]

Posted in Spectrometer, Usage | No Comments »

UPDATE ! : PNEUMATICS BACK TO NORMAL – ALL SYSTEMS OK

Sample handling is normal.  Please let the facility staff know or post a log book entry, if you notice anything out of the ordinary. NMR Facility Manager 9/18/15

Posted in Hardware, Spectrometer | No Comments »

ALERT ! – HOUSE AIR SUPPLY PROBLEM

We are aware of the drop in house air pressure that is leading to sample handling failure in all but the AV500 instrument.  The problem is identified to be a failed tubing connection to the buffer tank supplying all the three instruments.  Attempts are being made to fix this as soon as possible and updates […]

Posted in Hardware, Spectrometer, Usage | No Comments »

AV301 : UPDATE – SYSTEM USABLE BUT WORK NEEDS TO BE DONE WITH SHIMS

A damaged insert glass had caused problems leading to (a) in-correct sample landing (b) difficulty in shimming.   The broken glass itself makes shimming difficult but this problem is exacerbated greatly if the sample doesn’t land fully.   Users are reminded that before you go 60 mph,  make sure that the sample ‘Down’ button is lit up […]

Posted in Hardware, Spectrometer | No Comments »

ALERT ! – AV301 – UPDATE ON CURRENT CONDITION

With due apologies for the delay in this update, I would like to summarize the current condition of the instrument AV301 A problem with the 1H high power amplifier has been identified. We have come up with a temporary work around to make the instrument usable while a permanent fix is being worked out. The […]

Posted in Hardware, Spectrometer, Usage | No Comments »

UPDATE : NMR DATA SERVER FUNCTIONAL

Following the power outage over the weekend, the NMR data server phoenix has been restored to normalcy.  If you are still unable to access NMR data, please let us know.

Posted in Hardware, Spectrometer | No Comments »