ALERT ! : AV301 – PROBE CHANGE

The temporary double resonance probe will give way to the TXD-19F{1H,13C} probe and the instrument should be primed and ready for use on the morning of Wednesday, the 11th. Any change in this schedule will be posted here later.

Posted in Hardware, Spectrometer | No Comments »

AV301 : UPDATE – 1H{13C} probe as replacement

The 19F probe is temporarily replaced with a double resonance 1H {13C} probe. It is estimated to be three weeks before the 19F probe is active in AV301 again.  We will keep the user base updated of any change in this schedule.

Posted in Hardware, Spectrometer, Usage | No Comments »

AV300 : ALERT ! – INTERMITTENT LOCK PROBLEM

A sporadic malfunction in one of the circuit boards is causing the Lock system to behave unreliably, irrespective of the solvent you use.  A firmware reset of the board restores functionality.  The troubleshoot will be a prolonged operation due to the sporadic nature of the problem.  As of now, the Lock system is functional and […]

Posted in Hardware, Spectrometer | No Comments »

AV301 : ALERT ! LINESHAPE DETERIORATION

Preliminary observations indicate a possible damage to the insert glass inside the probehead. If it’s confirmed, we will replace the broken part  ASAP but, further shimming of the probe will be necessary to bring the lineshape to usable spec.  Look for further updates in this space.

Posted in Hardware, Spectrometer | No Comments »

AV500 – UPDATE : LOW VT STATUS

A low level liquid N2 sensor in the LN2 evaporator assembly is damaged.   We are trying to get this replacement part as soon as possible and bring the system’s low VT capability back on track.  Further updates will be posted when the status quo changes.

Posted in Hardware, Spectrometer | No Comments »

CORRECTION ! – LOW VT SYSTEM MALFUNCTION ON AV500 AND NOT DRX499

Sorry for the typo in my message earlier but the low VT system is down only in AV500 and not the DRX499. Please take note.

Posted in Hardware, Spectrometer | No Comments »

DRX499 : ALERT – LOW VT SYSTEM NON-FUNCTIONAL

Due to the liquid Nitrogen evaporator assembly malfunction, users cannot perform low T experiments that are below 0 C.  The hardware is being looked into and an update will be posted in due course.

Posted in Hardware, Spectrometer, Usage | No Comments »

DRX499 : ALERT – PROBE CHANGE TOMORROW

The currently installed BBI probe with a malfunction needs to be sent for repairs.  By tomorrow afternoon this probe will be pulled out and the backup triple resonance 1H-{13C, 15N} probe will be put in place and primed.  The replacement probe should be ready by noon  on Monday, the 28th Oct ’13.  Users are requested […]

Posted in Hardware, Spectrometer | No Comments »

ALERT : AV700 – PROBE CHANGE

The current solutions probe will give way to the 3.2 mm Solid State MAS probe on 28th October, 2013, Monday. Solution NMR users are requested to take note and plan their runs accordingly.

Posted in Hardware, Spectrometer | No Comments »

DRX499 : ALERT – BBI PROBE HANDICAP

Some of the capacitors inside the BBI probe are de-rating and this leads to the following behavior.  Whenever a gradient based experiment is performed, the pulse gradients affect the tuning and matching of the probe such that the wobb curve will be way off at the end of the experiment, leading to poor or no […]

Posted in Hardware, Spectrometer, Usage | No Comments »