INSTRUMENT UPDATES : AFTER THE POWER BREAKDOWN
We are all now aware of the power shutdown that occurred on Saturday last, towards evening, disrupting the NMR instruments, among other things. We are working diligently to revive all of them in as quick a time as possible and as we speak, only the AV500 is drawing continued attention. We appreciate the patience and understanding […]
Posted in Hardware, Software, Spectrometer, Usage | No Comments »
AV301 UPDATE : INSTRUMENT READY FOR USE (AGAIN…)
Users ! Please take a few moments to focus on sample insertion sequence, confirm that the sample is seated properly in the NMR magnet and then start zipping off with your experiments. Remember, one person’s haste can cost lot of lost instrument time for the entire user base. Thank you for your cooperation.
Posted in Hardware, Spectrometer, Usage | No Comments »
AV500 : INSTRUMENT READY FOR USE
Users can start booking time on this instrument. The probe is : 1H {19F, 13C} – Z gradient. Temperature range (without LN2 evaporator) : +10 C to +70 C. Standard Shim File : stdtxo
Posted in Hardware, Spectrometer, Usage | No Comments »
ALERT : AV301 DOWN DUE TO USER ERROR
A user has dropped a NMR sample tube without the spinner and lift air, into the magnet bore. As a result the spectrometer is not immediately useable. The issue will be resolved by Tuesday, April, 2nd and a follow up will be posted here.
Posted in Hardware, Spectrometer, Usage | No Comments »
AV800 – IMPORTANT INFORMATION
Users of 800 spectrometer are reminded that a flooding incident that occurred about a week ago damaged some of the sensitive electronic components, rendering the system unusable. Every effort is being put in, to revive the spectrometer as quickly as possible. Please look for further updates on this issue. Contact me if you need more […]
Posted in Hardware, Spectrometer | No Comments »
AV500 : PROBE CHANGE ALERT
The TXO-1H{19F,13C} Z-gradient probe will be installed in AV500 system on Friday, the 29th of March ’13. After priming the probe, it will be available for regular use by Monday, the 31st of March ’13. Users are alerted to plan accordingly.
Posted in Hardware, Spectrometer | No Comments »
AV301 – QUICK UPDATE
Since the 19F probe will take more time to fix, the automated BBI probe is being primed to replace the ailing probe. Along with line shape optimization, an existing VT system issue is also being addressed at this time. Owing to these, it has taken longer than expected to bring the instrument back on line. […]
Posted in Hardware, Spectrometer | No Comments »
AV301 : PROBE HEAD PROBLEM – SYSTEM DOWN
A possible breakage inside the probe makes the system unusable as of now. The matter is being looked into. An alternative solution will be provided as soon as feasible.
Posted in Hardware, Spectrometer, Usage | No Comments »
AV700 : PROBE CHANGE ALERT
The current 3.2 mm MAS solids probe will give way to the TXI-HCN triple resonance solutions probe soon. The solutions probe will be ready for use by Monday, 10th March. Users are requested make plans accordingly.
Posted in Hardware, Spectrometer | No Comments »
AV301 UPDATE : BROKEN PARTS IN PROBEHEAD
Most probably due to a combination of poor quality NMR tubes that were spun and a less than careful sample positioning in the sample spinner, by various users, the expensive quartz insert glass tubes inside the probe have been shattered. All the woes of the past few days, with the sample seating, difficulty in spinning […]
Posted in Hardware, Spectrometer, Usage | No Comments »