DRX499 : ALERT – Lineshape Optimization in progress
Following the feedback from several users about asymmetric lineshape distortion in spectra, the lineshape optimization routine is setup on the instrument, which will run overnight. If the problem persists after this, other possible causes to this issue will be explored. Users who have blocked the instrument today are requested to move to other instruments for […]
Posted in Hardware, Software, Spectrometer | No Comments »
AV700 : INSTRUMENT IS FUNCTIONAL
AV700 is fully functional and will be hosting a solids MAS probe currently. Probe change to a solutions probe will be announced here subsequently.
Posted in Hardware, Spectrometer | No Comments »
AV800: system down due to gradient malfunction
We are waiting for a hardware replacement part. The expected timeline for the same is end of this week i.e. 8 Feb 2013. Any changes to this schedule will be posted here.
Posted in Hardware, Spectrometer | No Comments »
AV700: UPDATE
A pneumatic control board failed in the AV700 console, making the spectrometer unusable. We have been waiting for a replacement of the said board from the manufacturer. The current status is that, this board should arrive at the facility in roughly two weeks time. Once the system is up and functioning, I will post another […]
Posted in Hardware, Spectrometer | No Comments »
AV301 : UPDATE ON VT
The chiller unit is malfunctioning and therefore variable temperature is not available on the instrument until further updates. All spectra need to be recorded at 298 K i.e. room temperature.
Posted in Hardware, Spectrometer | No Comments »
AV301 : PROBE UPDATE – 19F probe back
The erstwhile damaged 19F probe has been repaired in house and has been tested to function satisfactorily. We thank the users for their patience and understanding while waiting for the 19F probe to be available. Please let me or Paul know if you observe anything of significance with the TXD 1H {19F,13C} probe.
Posted in Hardware, Spectrometer | No Comments »
AV301 UPDATE : NEW AUTOMATED BROADBAND PROBE
The 19F probe has been compromised and will need extensive repairs. In place of the old probe, the new BBI probe is installed in the system. IMPORTANT NOTE : This is a new generation probehead that does NOT allow you to do manual tune/match i.e. WOBB. You must use the command ‘atma’ instead and the […]
Posted in Hardware, Software, Spectrometer | No Comments »
AV301 : ALERT – 19F CHANNEL MALFUNCTION
A probe head related issue is being looked into . If it cannot be fixed within the next hour, an alternative will be provided. Please wait for further updates.
Posted in Hardware, Spectrometer | No Comments »
AV500 : PLANNED PROBE CHANGE and B0 FIELD CHANGE
The X{1H} -Z grad BBO probe will be tested and installed on 26th Oct’12, Friday. The probe optimization will be performed over the weekend and the probe should be ready for use by Monday, the 28th of Oct ’12. Along with this operation, the B0 field will be readjusted to re-position the LOCK field to […]
Posted in Hardware, Software, Spectrometer, Usage | No Comments »
AV301 : SAMPLE SPINNING ISSUE
A possible blockage in the pneumatic system is causing the sample not to spin correctly. Related to this is the fact that, the BSMS keyboard might indicate the sample as ‘missing’, even when the sample is seated inside. The shim stack needs to be removed to clear this problem. Users are requested not to use […]
Posted in Hardware | No Comments »