ALERT ! AV500 OK – WITH IMPORTANT CHANGE – PLEASE READ !
CURRENT STATUS : First I will highlight that the sporadic Topshim problem is gone now and the shimming is behaving as expected. That should come as a relief to all of you who exclusively use 2H lock solvents other than (H2O+D2O) or (CH3OH+D2O). IMPORTANT – PROBEHEAD CHANGED TO BBI In the process of system troubleshoot, […]
Posted in Hardware, Spectrometer, Usage | No Comments »
ALERT ! AV500 – TOPSHIM OCCASIONAL ISSUES
Since the majority of lock solvents used for routine NMR on this instrument are fully dueterated, for those users, the Topshim routine heavily relies on the ability to get good 2H NMR signal response from the hardware. Following reports of sporadic failure of Topshim, I have done a bit of deeper digging and there is […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE : AV500 – SAMPLE HANDLING IS FINE – NO PROBLEMS DETECTED
This is the second instance where I saw a post saying that sample couldn’t be ejected. But I didn’t detect any issues per se and all the vitals of the instrument are normal. If users need some clarification and refresher as to how the sample eject/inject sequence works and what are the possible ways one […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! : AV500 – SAMPLE HANDLING IS NORMAL
Sample eject anomaly reported by folks today morning is resolved now. Along with it, I have also checked the general health of the instrument and all vital signs are normal. NMR Facility Manager 6/8/2021
Posted in Hardware, Spectrometer | No Comments »
UPDATE : AV500 – IS BEHAVING NORMALLY
In continuation of today morning’s instrument log post, I can confirm that the ‘wobb’ function as well as the ‘Topshim’ functionality works just fine. I will also add that the probehead is tuning as expected as well for the 1H channel. I have also run some test runs and the overall behavior of the probe […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE : AV301 – SAMPLE HANDLING ERROR FIXED
The sample eject/inject issue has been resolved. In the process of checking other BSMS related functions, I have also updated the optimal default shim file : stdpabbi. When you are in doubt, always use : rsh stdpabbi to take advantage of the best higher order shims that will give you the best lineshape for all […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE : DRX499 – IS BACK TO NORMAL
The acquisition error reported in the log is fixed now. NMR Facility Manager 3/29/2021
Posted in Hardware, Spectrometer | No Comments »
MNOVA : UPDATE ON THAT IMPORTANT UPDATE ! PLEASE READ (JUST ONCE MORE !…)
Hi Users Here is a Happy Exception to the Caveat I transmitted moments ago : The software needs to validate its license by connecting to the License Server on Campus only ONCE IN 90 DAYS. This means that, you are off the hook in connecting to the Husky OnNet for 90 days, once you check […]
Posted in Software, Usage | No Comments »
IMPORTANT UPDATE ABOUT THE MNOVA SOFTWARE USAGE ! PLEASE READ !
Dear Users You must remember this very important caveat to use the MNova you downloaded : You must be either on Campus or must have connected your computer to the Husky OnNet VPN network when working from home for the software to run. The licenses are validated/protected to avoid unauthorized use by ensuring that the […]
Posted in Software, Usage | No Comments »
ANNOUNCEMENT ! – VERSATILE MNOVA DATA PROCESSING SOFTWARE AVAILABLE FOR USE !
Dear NMR Community of UW I am excited to announce that the industry leading MNOVA (a.k.a. MestreNova for oldtimers) NMR data Processing Software is now available free to you. Student Technlogy Fee funded initiative Thanks to the generous support provided by the Student Technology Fee (STF) resource of UW, MNOVA NMR software is now available […]
Posted in Software, Spectrometer, Uncategorized | No Comments »