NOTE ! : DRX499 – (HOPEFULLY) USEFUL EMBELLISHMENT
Hello after a long break and Happy Wednesday and Happy New Year 2020 ! DRX499 Since we all still have to tune/match the TXO-19F,13C {1H} NMR probehead by hand by crouching under the magnet, we are throwing some light on the process. We have provided a corded LED lamp to light up the probehead space under […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV300 : FROZEN TOPSPIN SCREEN REMEDIED
The cause of the frozen screen upon login is this : When there is no dataset that is already open in the spectrum display area (blue tile), we usually drag and drop an existing dataset into it. For some reason that has to do with the way the Display system and Topspin interact, this causes […]
Posted in Software, Spectrometer, Usage | No Comments »
UPDATE ! DRX499 – SYSTEM BACK TO NORMAL
Acquisition hardware issue has been resolved. System tested and back to normal. Please let me know, if it is otherwise. NMR Facility Manager 10/8/2019
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV500 – SYSTEM BACK TO NORMAL OPERATIONS
There are two updates to be highlighted here : AV500 is back to a three channel configuration after an (expensive) replacement of a failed circuit board. You can therefore plan {1H, X, Y} type experiments. The TBI-1H {13C,X} – Z gradient probe has been placed back in the magnet. This probe has a dedicated 13C […]
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE ! AV500 – INTERIM SOLUTION IN PLACE
While we are waiting for replacement of the compromised circuit board, we have patched this unit together to function as a two-channel instrument. The BBI-Z gradient probe is paired with this so that you can pulse two nuclei at a time i.e. 13C with 1H decoupling OR 31P with 1H decoupling, as examples. In a […]
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE : AV500 – SYSTEM WILL BE DOWN UNTIL FURTHER NOTICE
The RF related issue is not fully resolved yet. As I mentioned before, there is a potential probehead change that might be called for. There is no clear correlation between the electronics problem mentioned above and the probehead’s condition at this point. This probe has been giving a rather noisy ‘wobb’ signal as many of […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE – 2 : AV300 IS BACK ON TRACK !
The AV500 is the only system that is yet to be revived. But, I would like to alert you to the fact that, the revival of this instrument might take longer. In fact, it is possible that it is not ready by today. More updates will be coming as the situation evolves. Thank you, as […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! – AV300 AND DRX499 ARE UP; OTHER SYSTEMS STILL DOWN
AV301 : A computer related glitch is the main cause behind the failure and our IT Tech is working on this. AV500 : A malfunctioning circuit board is the obstacle to reviving the system fully. It is possible that the probehead is having issues too. Stay tuned for more updates. NMR Facility Manager 9/9/19
Posted in Hardware, Spectrometer | No Comments »
ALERT ! SPORADIC ISSUES WITH NMR SYSTEMS – Thunder and Lightning !
The excitement over the weekend brought by the Weather gods, did leave a bit of an imprint on the NMR systems. Though you may not see anything amiss immediately, some of them may not work as expected. As I speak, I am making sure the systems get back to normal, one by one. I will […]
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV700 – SOLUTIONS TO SOLIDS SWITCH SOON
All 700 NMR users please take note that the system will move to solid state NMR mode roughly in a week’s time from today. Our current estimate is that the instrument will remain in solids mode for about 2 months. We shall assess the situation towards the end of that window and see if the […]
Posted in Hardware, Spectrometer | No Comments »