AV300 ALERT! – PROBE SWAP NOW
Due to the frequently recurring 1H wobb issue, which has occurred again today, the BBI probe is being pulled and sent for repair. The new probe will be the TXD {1H, 19F, 13C} probe typically seen in the av301. Thank you for your patience. Please contact the NMR Facility Managers with any questions or concerns.
Posted in Hardware, Spectrometer | No Comments »
AV300 – UPDATE ! – SYSTEM USABLE AGAIN
We could fix the tuning problem of 1H in house. But, I cannot emphasize enough, the following : BE EXTREMELY GENTLE WITH THE TUNE AND MATCH RODS. It is only a question of time before the problem recurs. Plans are in the pipeline to provide a backup probehead for AV300 and service this probehead. […]
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV300 – 1H CHANNEL OF PROBEHEAD NON-FUNCTIONAL
The 1H channel doesn’t tune as usual. We will try to provide a workaround as soon as we can. Please stay tuned ! NMR Facility Manager 11/2/15
Posted in Hardware, Spectrometer | No Comments »
AV500 UPDATE – TXI-HCN PROBE – Z GRADIENT – TOPSHIM READY
We will announce once the probe is calibrated for its first use with all relevant parameters optimized. Thank you for your patience. NMR Facility Manager 11/23/15
Posted in Hardware, Spectrometer | No Comments »
AV500 UPDATE – TXI-HCN PROBE – Z GRADIENT – TOPSHIM READY
We will announce once the probe is calibrated for its first use with all relevant parameters optimized. Thank you for your patience.
Posted in Hardware, Spectrometer | No Comments »
UPDATE : AV300 RETURNED TO USE
After attending to the tune issue on the probe that is directly caused by users using fingers to turn the turn rods, push past the stop point of the tune rod, or using the tune tool incorrectly, the av300 BBI probe has been repaired and is returned for use. The NMR Facility cannot stress enough […]
Posted in Hardware, Spectrometer | No Comments »
ALERT: AV300 UNABLE TO RUN DUE TO TUNE ISSUES
Users: The av300 BBI probe has been misused again and is no longer tuning/matching on the 1H, which precludes proton and any 1H-decoupled-X-observe NMR. Until the NMR Managers can take a look, please do not run on the av300 because using the av300 in its current state risks further damage to the console. Please watch […]
Posted in Hardware, Spectrometer | No Comments »
LOGIN ISSUES RESOLVED on AV300 and DRX499
Login issues have been resolved on both the av300 and drx499 spectrometers.
Posted in Software, Spectrometer | No Comments »
ALERT ! : DRX499 – UWNETID LOGIN DOWN…TEMPORARILY
An issue similar to that of AV300 is not allowing users to login with their NetID currently. We are working on this and post an update as soon as the issue is resolved.
Posted in Software, Spectrometer | No Comments »
AV300 : USER LOGIN NOT WORKING
The UWNetid is not functional and we are working on this. I will provide an alternative or any useful updates on the situation, shortly. NMR Facility Manager 11/16/15
Posted in Software, Spectrometer | No Comments »