UW CHEMISTRY NMR
Announcments & Equipment Logs
UPDATE ! AV300 – PENDING ISSUES RESOLVED – SEE DETAILS
First of, the system is ready for use ! NETWORKING ISSUE : The temporary network glitch is resolved now and users can login with their UWNetid and store data in the UDRIVE. If you notice anything different with your account, please send me an email. SAMPLE LANDING AND SPINNING OK : Lot of debris and […]
Posted in Hardware, Spectrometer | No Comments
UPDATE ! AV301 – SAMPLE HANDLING AND LOCK WORKING OK
The issue of BSMS not responding, leading to sample handling and lock was a temporary issue, probably caused by a voltage surge. The issue is resolved and the system is functioning normally now. NMR Facility Manager 2/12/2019
Posted in Hardware, Spectrometer | No Comments
UPDATE ! AV500 – THE TBI PROBE IS BACK
The mechanical issue compromising the 1H match has been fixed in-house and the probehead is back in the magnet. We have replaced the broken 1H match knob with a homemade knob. I will show a picture below that clearly identifies the new homemade knob. Only for 1H Match, you need to use the following tune/match tool […]
Posted in Hardware, Spectrometer | No Comments
ALERT ! AV300 – SAMPLE LANDING ISSUE – UPDATE
Sample spinning with this instrument is not working and this is also related to the BSMS system’s observation that sample is ‘missing‘. We are looking into most plausible causes before we decide to pull the probe out, to minimize the down time with the instrument. Please note that the only alternate probehead that can […]
Posted in Hardware, Spectrometer | No Comments
ALERT ! AV500 – PROBE CHANGED TO HCP-Z gradient
We discovered a broken part in the TBI probe that made the 1H matching process untenable. While we are working on this, we have placed the triple resonance 1H{13C, 31P}-Zgradient probe in the magnet, as a temporary measure (hopefully). We have every intention of returning to the excellent TBI probe once it is ready to […]
Posted in Hardware, Spectrometer | No Comments
ALERT ! AV500 – 1H MATCHING ROD MAY NEED FIXING
We are aware of reports from some users that the 1H matching rod is still not moving smoothly. We will get to this as soon as possible and post an update of the results. NMR Facility Manager 1/29/2019
Posted in Hardware, Spectrometer | No Comments
UPDATE ! AV500 – SYSTEM READY FOR USE
The probe mechanics are fixed to normalcy. Please go ahead and use the instrument. Do not hesitate to post in the log book if you encounter difficulty. The tuning/matching rods, are near the magnet legs. They are no longer hanging from underneath the probe. This will prevent accidental loosening of the screws in the probehead […]
Posted in Hardware, Spectrometer | No Comments