UPDATE ! BOTH THE 300 NMR-s ARE BACK TO NORMAL

Please go ahead and reserve time and start using the instrument. Let us know if you have any issues. NMR Facility Manager 1/23/18

Posted in Hardware, Spectrometer, Usage | No Comments

UPDATE ! NMR SYSTEMS LIMPING BACK TO NORMAL

AV500 and DRX499 are up and running.  We are still having issues with AV301 and AV300.  Until they are resolved, the reservation calendar for these instruments will be locked.  Please look for my next update announcing normalcy of these two instruments and then re-book your slots. Currently reserved slots on these instruments have been cleared […]

Posted in Spectrometer, Usage | No Comments

ALERT ! FULL REVIVAL OF NMR SYSTEMS STILL IN PROGRESS..PLEASE BE PATIENT

Following the building wide power shutdown, several networking components are still in the process of being revived.  This translates to extremely slow or unresponsive NMR host computers and networked peripherals.  IT staff are working busily addressing this. We will post an update once all systems are normal I will exploit this situation to highlight  why […]

Posted in Host Computer, Operating System | No Comments

ALERT ! ALL NMR SYSTEMS WILL BE POWERED DOWN – Bagley Electrical Shutdown on Tuesday, the 23rd

Approximately by 4 pm today, we will shut down all the NMR systems in Bagley Hall, in preparation for the building wide Electrical Reboot that Eric Strakbein announced for tomorrow.  Erik’s mail says that it will be a 2 hour shutdown beginning at 5 am.  We will revive the facility as soon as we can […]

Posted in Spectrometer, Usage | No Comments

ALERT ! DRX499- IMMINENT PROBE CHANGE – Wednesday next week

The freshly serviced TXO – 19F,13C {1H}-Z Gradient probe will be installed in DRX499 on Wednesday, the 24th of January ’18.  It is our hope that the system will be primed and ready before the end of the day. I will update the post if any new information needs to be conveyed. NMR Facility Manager […]

Posted in Hardware, Spectrometer | No Comments

DRX499 : SYSTEM DOWN DUE TO HARDWARE FAILURE

Both the power amplifiers have damaged power supplies, which are being repaired in house. It is hoped that it will be a quick fix. The instrument will be released as soon as one of the amplifier is tested to be ok. NMR Facility Manager 12/15/17

Posted in Hardware, Spectrometer | No Comments

UPDATE : AV300 – ATMA AND ATMM WORKING NORMAL – AS OF NOW…

The noise noted by users on WOBB curve while ATMA is fired up, seems to be an intermittent issue and it is not observed as of now.  Also, the spectra themselves didn’t show any spurious noise spikes even when the spikes on WOBB curve was observed. We will keep watching for this behavior to return […]

Posted in Hardware, Spectrometer | No Comments

UPDATE : DRX499 – SYSTEM RELEASED FOR ROUTINE USE

Please use the starting shim file : stdbbo before you touch up lower order shims.  If you need any specific help with shimming, please contact either Adrienne or myself. NMR Facility Manager 12/1/17

Posted in Hardware, Spectrometer, Usage | No Comments

UPDATE : AV300 AND AV301 – THINGS ARE BACK TO NORMAL

The ATMA errors as well as communication issues between the PC and console seemed to have stemmed from a possible power disruption, that was momentary.   Both the systems working normally as I write this.  If you still notice any issues, please post a message or send an email. NMR Facility Manager 11/29/2017

Posted in Hardware, Spectrometer | No Comments

ALERT ! DRX499 – IMMINENT PROBE SWAP

The TXO – 19F,13C {1H} probe needs to be refurbished to its full potential and this is long overdue.  We will therefore replace this probe with our backup BBO{1H}-Z gradient probe on Thursday, the 30th of November.   With the availability of Gradient shimming, the work up of backup probe should be quicker and we will […]

Posted in Hardware, Spectrometer | No Comments