UPDATE ! AV500 — SYSTEM IS FULLY OPERATIONAL
Users can start booking time on the instrument immediately. The probehead is the same as before i.e. BBI – Z-gradient. The issue with this probe has been resolved. Topshim has been tested and working correctly both in 1D and 3D modes. Please contact Adrienne or myself, if you have any questions. Sorry of this long […]
Posted in Hardware, Spectrometer, Usage | No Comments
ALERT ! AV500 – SYSTEM DOWN !…
Being in Seattle, I am sure you can identify, when I say “When it rains it pours ! “ Right after the replacement of the faulty circuit boards, we had a short power disruption yesterday and it took some reconfiguration in the console to bring things to normal. Today, I have found the BBI probehead, […]
Posted in Hardware, Spectrometer | No Comments
ALERT ! AV500 – SYSTEM FUNCTIONAL WITH A POSSIBLE CAVEAT
The power fluctuation yesterday left a few issues lurking in the console, which are resolved now. The X-channel is working as expected but the 1H channel still has a detuning problem. Due to this you will notice rather noisy 1H spectrum – this is related to the wobb curve simply moving to a detuned state […]
Posted in Uncategorized | No Comments
UPDATE – AV301 : VT SYSTEM IN STABLE MODE
As mentioned, we have isolated the chiller that was freezing up samples when the probe heater was turning off. You can collect data at room temperature and above currently. We will update you when the low T capability is restored.
Posted in Hardware, Spectrometer | No Comments
ALERT AV500 – 1H WOBB INOPERABLE
After the short power outage at approximately 3 pm on March 1, the av500 is not communicating with one of the new boards. NMR Facility managers will take a look at it first thing Wednesday morning. In the meantime, please reschedule your av500 time to another instrument.
Posted in Hardware, Spectrometer | No Comments
NOTE – AV301 : TEMPERATURE ISSUES
We are aware of the issue noticed by most of you where the VT heater simply shuts itself off, leaving the cold air in the probe which might even freeze some high freezing point solvents such as DMSO. Since there are more than one possible cause, that might involve the ATMA Probe also, we are […]
Posted in Hardware, Spectrometer | No Comments
NOTE ! AV500 – LOCK IS WORKING OK
The lock problem noticed by users is because of a non-optimal value of lock phase. Normally the AutoLock routine tries to adjust this value within a range but once it falls outside of this, lock fails and manual intervention is necessary. That is done now and everything should be normal. Please note that Topshim on […]
Posted in Hardware, Spectrometer, Usage | No Comments
UPDATE ! AV500 – FAULTY CIRCUIT BOARD REPLACED
One of two boards have been replaced and we are waiting for the second board to arrive. But, with this replacement, full functionality is restored to the most part. The 1H 90 pulse width is now 7.5 us at the usually employed lower power (contrast with the 18 us we reported a few days ago) […]
Posted in Hardware, Spectrometer, Uncategorized | No Comments
UPDATE ! AV500 – INSTRUMENT USABLE WITH LIMITED FUNCTIONALITY
We have ordered the necessary replacement parts and asked the manufacturer to ship them asap. I am hoping that it will be Thursday before we get these parts. I am releasing the instrument for nominal use, with the following caveats, until we get the new boards in : The wobb signals WILL be noisy, but […]
Posted in Hardware, Spectrometer | No Comments
ALERT ! AV500 – HARDWARE ISSUE
One of the RF circuit boards is malfunctioning and the instrument is not ready for immediate use. Those of you who have blocked time are requested to cancel the same for the day. We will post an update about the progress soon. NMR Facility Manager 2/23/16
Posted in Hardware, Spectrometer | No Comments