ALERT ! AV300 : WOBB ISSUE BEING INVESTIGATED
The ‘wobbly’ wobb has been an ongoing issue that we are aware of and haven’t gotten to with full force. We will be doing the same today and try to understand what’s causing it. Based on observations made thus far, it might take a while before we fix this in house or get some parts […]
Posted in Hardware, Spectrometer | No Comments
AV300: BBI PROBE INSTALLED
The BBI 1H{X} probe has been installed on the av300. It is released for use. We are still observing some noise on the wobb for both channels, which we are working on resolving. Thank you for your patience.
Posted in Hardware, Spectrometer, Usage | No Comments
USERS NOTE ! AV301 AND OTHER INSTRUMENTS – NEED YOUR HELP
Those of you who change the temperature to anything other than the room temperature i.e. 298 K are requested to be considerate to the next user. Please change the default temperature to 298K before you logout. Inadvertently, you might be freezing the next user’s sample or killing the molecule that may not be cozy at […]
Posted in Spectrometer, Usage | No Comments
DRX499 UPDATE – TXO PROBE INSTALLED
To all users expecting a probe swap on the drx499, the TXO 1H {13C, 19F} probe is currently installed.
Posted in Hardware | No Comments
AV300 ALERT! – PROBE SWAP NOW
Due to the frequently recurring 1H wobb issue, which has occurred again today, the BBI probe is being pulled and sent for repair. The new probe will be the TXD {1H, 19F, 13C} probe typically seen in the av301. Thank you for your patience. Please contact the NMR Facility Managers with any questions or concerns.
Posted in Hardware, Spectrometer | No Comments
AV300 – UPDATE ! – SYSTEM USABLE AGAIN
We could fix the tuning problem of 1H in house. But, I cannot emphasize enough, the following : BE EXTREMELY GENTLE WITH THE TUNE AND MATCH RODS. It is only a question of time before the problem recurs. Plans are in the pipeline to provide a backup probehead for AV300 and service this probehead. […]
Posted in Hardware, Spectrometer | No Comments
ALERT ! AV300 – 1H CHANNEL OF PROBEHEAD NON-FUNCTIONAL
The 1H channel doesn’t tune as usual. We will try to provide a workaround as soon as we can. Please stay tuned ! NMR Facility Manager 11/2/15
Posted in Hardware, Spectrometer | No Comments
AV500 UPDATE – TXI-HCN PROBE – Z GRADIENT – TOPSHIM READY
We will announce once the probe is calibrated for its first use with all relevant parameters optimized. Thank you for your patience. NMR Facility Manager 11/23/15
Posted in Hardware, Spectrometer | No Comments
AV500 UPDATE – TXI-HCN PROBE – Z GRADIENT – TOPSHIM READY
We will announce once the probe is calibrated for its first use with all relevant parameters optimized. Thank you for your patience.
Posted in Hardware, Spectrometer | No Comments
UPDATE : AV300 RETURNED TO USE
After attending to the tune issue on the probe that is directly caused by users using fingers to turn the turn rods, push past the stop point of the tune rod, or using the tune tool incorrectly, the av300 BBI probe has been repaired and is returned for use. The NMR Facility cannot stress enough […]
Posted in Hardware, Spectrometer | No Comments