UW CHEMISTRY NMR
Announcments & Equipment Logs
Spectrometer
ALERT ! NMR ACCESS AND BAGLEY ELECTRICAL SHUT DOWN ON 29TH OCT ’17
TO ALL NMR FACILITY USERS – PLEASE READ ! I am sure all of you have seen the announcement from Erik Strakbein about the Bagley Hall Electrical Shutdown that is scheduled over coming weekend. In anticipation, we will be preemptively shutting down ALL the NMR SYSTEMS before the electrical outage happens. According to current plans, […]
Posted in Hardware, Spectrometer, Uncategorized | No Comments »
UPDATE ! DRX499 – HARDWARE UPGRADE IN PROGRESS
As some of you might have noticed, the Hardware Upgrade you saw in the reservation calendar pertains to upgrading the system with Gradient Shimming and 2H observe capability, that hitherto was not present. Although we had an old but functional gradient amplifier, the console was not capable of utilizing it to do gradient shimming, which […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV500 – PROBE CHANGED TO TRIPLE RESONANCE – HCP
The BBI probe’s VT dewar is faulty, which most probably contributed to an over temperature event. Since we have to wait for the replacement part to arrive (might be a couple of weeks), the following probehead has been placed and optimized for regular use : TXI – 1H {13C, 31P) – XYZ gradient This probe […]
Posted in Hardware, Spectrometer | No Comments »
UPDATE ! AV301 : SYSTEM IS BACK IN ACTION
Pneumatics system is fixed and working properly. Ready for regular use. NMR Facility Manager 9/26/17
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV500 – SYSTEM DOWN DUE TO VT MALFUNCTION
A runaway VT system has overheated the sample space and the system needs time to recover to normal temperature range. We are also investigating the cause of this. Until further notice, the system is not available for regular use. Updates will be posted as and when we get more information to report. Thank you for […]
Posted in Hardware, Spectrometer | No Comments »
ALERT ! AV301 : PNEUMATICS MALFUNCTION
Sample Eject and Spin is not functioning and initial troubleshoot points to a problem within the Pneumatics controller. Until the unit is revived the spectrometer cannot be used. Please stay tuned for further updates. NMR Facility Manager 9/25/17
Posted in Hardware, Spectrometer | No Comments »
UPDATE : AV300 : BBO MANUAL PROBE OPERATIONAL – this is a back up probe
While we are fixing the ATMA probe, we have provided a surrogate broadband probe in AV300 for routine runs. Please note that, the default shims provided may not be the best to just run spectra out of the box. You have to touch up the low order shims by hand to get decent spectra. We […]
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE : AV300 : ALTERNATE PROBE IS BEING READIED
We are working with the manual BBO probe to make it optimal for regular use. In parallel, the ATMA probe repair will be expedited so that we can get back to automated operation with minimal disruption to that routine. Thank you for your understanding NMR Facility Manager 9/21/2017
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE : AV301 : VT CONTROL UPTO 0 DEG C POSSIBLE NOW.
Some of you might have noticed that the sample temperature cannot be lowered below ambient on AV301. The heat exchanger a.k.a. chiller that makes this possible, is now serviced and working as expected. This gives you the choice to run lower than RT experiments upto 273 K max. This implies that you must set the […]
Posted in Hardware, Spectrometer, Usage | No Comments »
UPDATE ! AV500 – SYSTEM BACK TO NORMAL
All functioning as expected. Please give us a feedback, when you notice abnormal behavior. Thank you NMR Facility Manager 9/18/2017
Posted in Hardware, Spectrometer | No Comments »