ALERT – DRX499 : WALK ON TIMING UPDATE

To bring the two 500 MHz instruments on par and to ease the load on the AV500 spectrometer, that is being heavily booked of late,  users are informed that the walk-on duration on DRX499 is extended to a 3 hour limit from the current 1 hour limit. This should help users significantly. For instance one […]

Posted in Spectrometer, Usage | No Comments »

AV301 – UPDATE : LINESHAPE DISTORTION

We are aware of this issue and looking into it. Preliminary investigation has ruled out the probehead to be the immediate cause and we are in the process of testing other components.  In the meanwhile,  users are encouraged to use the pulse program zg30 in place of the usual zg experiment to record the spectra.  […]

Posted in Hardware, Spectrometer | No Comments »

ALERT – AV500 : DEUTERIUM GRADIENT SHIMMING NON-FUNCTIONAL

A hardware board malfunction makes the 2H Topshim non-functional.  Repair attempts are in progress with the faulty circuit board.  Users can still use Topshim 1D and 3D for proteo-solvents and aqueous sample preps.  Update will be posted in due course.

Posted in Hardware, Spectrometer | No Comments »

UPDATE : ALL FACILITY INSTRUMENTS ARE OPERATIONAL

All instruments tested to be stable and in normal operational mode.  The FTP server is hosting the data trees of all instruments, as expected.

Posted in Hardware, Spectrometer | No Comments »

ALERT ! FACILITY POWER SHUTDOWN TOMORROW, Sept. 5th

As announced earlier, the NMR facility instruments will be shutdown by 3.30 pm tomorrow in anticipation of a scheduled power shut down to the NMR lab i.e. Room:37, Bagley Hall by early morning on Saturday the 6th September. The power shut down is announced to last until 3.00 pm on Saturday. The instruments will be […]

Posted in Hardware, Spectrometer | No Comments »

EARLY ALERT : POWER SHUTDOWN NEXT WEEK – Friday, Sep. 5th

Users might have noticed the power disruption announced for Bagley Hall on Sep. 6th, Saturday.  This planned event will affect the NMR lab Rm:37 from the early morning of Saturday 6th until about 3.00 pm on the same day.   The facility instruments, therefore will be shutdown by the afternoon of Friday the 5th Sep.  and […]

Posted in Hardware, Spectrometer | No Comments »

AV 700 UPDATE: Returned to use with TXI probe

The problems with the av700 pneumatics board have been resolved. The AV700 is available for use with the solution TXI probe. The av700 will be changing to solids in a couple weeks.

Posted in Spectrometer, Usage | No Comments »

AV700 – UPDATE : PNEUMATICS BOARD MALFUNCTION

Probably due to a voltage surge this weekend past,  the pneumatics control board on 700 appears compromised. No VT regulation or sample transport is possible due to this and we are energetically trying to resolve the issue.  Please look for further updates in this space. The machine is unusable until then.

Posted in Hardware, Spectrometer | No Comments »

DRX499 : UPDATE – OPERATIONAL WITH LIMITED FUNCTIONALITY

The spectrometer with the TXO probe is operational fully with 1H and 13C channels. The 19F is not yet ready for prime use as I write this and I will post an update once that issue is resolved as well. The instrument is open for regular use, with the above caveat.

Posted in Hardware, Spectrometer, Usage | No Comments »

UPDATE: AV300, AV301, AND AV500 AVAILABLE FOR USE

The errors being seen on the av300, av301, and av500 from the power fluctuations yesterday have been resolved. The spectrometers are now available for use. Please contact the facility managers and use the logbooks should you find further problems. Thanks.

Posted in Usage | No Comments »