ALERT ! : PROBE CHANGE – AV500 and DRX499

Probe Swap date :  August 1st, Friday AV500 : The existing TXO probe will be replaced by the Z-gradient BBI probe and primed for use.  The system should be ready for use before the day is over. DRX499 : The TXO-19F probe will take the place of the current probe. Due to unavailability of gradient […]

Posted in Hardware, Spectrometer | 1 Comment »

UPDATE – AV301 : SYSTEM IS UP !

Users can reserve time as usual and start using the instrument. Do keep us posted if you see anything abnormal.

Posted in Host Computer, Operating System, Spectrometer, Usage | No Comments »

AV500 – SHIM UPDATE

Some users had complained that they are not able to get good line shape for their sample and Topshim software did not help, apparently, either.   My experience today has been different. The spectrum shown below is the result of a successful shimming operation with Topshim, for running this reasonably demanding 1D spectrum that suppresses 95% […]

Posted in Spectrometer, Usage | No Comments »

UPDATE ! DRX499 – SYSTEM BACK TO NORMAL

Please let us know if you face any issues again.

Posted in Hardware, Spectrometer | No Comments »

ALERT – DRX499 – COMMUNICATION SOFTWARE BEING REINSTALLED

Normal acquisition is not proceeding due to a corruption of low level software on the host computer.  This is being re-installed and the system should be up soon.

Posted in Software, Spectrometer | No Comments »

UPDATE – AV700 : ISSUES RESOLVED

Acquisition as well as Lock problems reported earlier have been fixed. Please send us a feedback if you still notice any issues with as much information that you can provide.

Posted in Hardware, Spectrometer | No Comments »

UPDATE : DRX499 – AVAILABLE FOR USE

Many of the issues on the drx499 users have noticed in the past two weeks were due to a faulty board. This board has been replaced and the console is now fully functional. A test run shows normal operation. Please contact Adrienne (amr0311@uw.edu) or Rajan (paranji@chem.washington.edu) with questions.

Tags:

Posted in Hardware, Spectrometer | No Comments »

UPDATE : DRX499 – AWAITING REPLACEMENT CIRCUIT BOARD

A faulty ADC board is recognized to be the most probable cause of problems and an order for its replacement was placed with instructions for expedited shipping. We will update the status once the board arrives and we test it out in the system.

Posted in Hardware, Spectrometer | No Comments »

ALERT ! : DRX499 – SYSTEM DOWN – HARDWARE MALFUNCTION

The receiver amplifier is suspect and is giving rise to glitches along the spectral range. Some users had already reported the ‘wobb’ curve showing peculiar noisy features that is also traced to this cause.  The replacement/repair options of the board are being explored and we will try to fix this as soon as feasible.  Until […]

Posted in Hardware, Spectrometer | No Comments »

AV301 : NOTE On PROBE CHANGE

Following the stuck sample tube incident, a troubleshooting attempt of the ATMA – PABBI probe led to an unfortunate damage to the probe and it is being sent to Bruker for repairs. This is the reason behind the hurried replacement in this instrument, with the TXO-19F {1H,13C} triple resonance probe.  If you have further questions […]

Posted in Hardware, Spectrometer | No Comments »