AV301 – UPDATE : GRADIENT UPGRADE
The hardware installation is complete but there are few more configuration steps that involves the software. I hope that this can be completed by tomorrow, unless some issue crops up in the meanwhile. I will keep you posted. NMR Facility Manager 11/3/15
Posted in Hardware, Spectrometer | No Comments
ATTENTION – AV301 USERS ! – HARDWARE UPGRADE – GRADIENT CAPABILITY IS COMING TO AV301
We are excited to announce that the AV301 instrument will get new hardware upgrade that will make it the third instrument in Bagley 37 to perform gradient assisted spectroscopy. In combination with the ATMA-BBI probe, this will become a fully automated spectrometer for routine spectroscopy. Other than injecting the sample, every other operation can be done from […]
Posted in Hardware, Software, Spectrometer | No Comments
UPDATE – DRX499 – INSTRUMENT RELEASED FOR USE WITH TXO PROBE
The BBO probe has been found to have multiple hardware issues which are being addressed. To get to back to productive runs, we have placed the earlier TXO – 19F,13C {1H} – Z gradient probe for regular use. I would like to point out that the non-spinning shims are not upto the best spec but […]
Posted in Hardware, Spectrometer, Usage | No Comments
ALERT ! – DRX499 – PROBE WOBB MALFUNCTION – X CHANNEL
Tuning for all X-band nuclei i.e. 13C, 31P, 11B, etc. is not possible at this time, though the 1H channel is functioning fine. We are looking into this and post an update as we make progress.
Posted in Hardware, Spectrometer | No Comments
UPDATE – AV500 IS OK !
Sample transport is working normally. As a related remark, if you are using special NMR tubes, that have a shifted center of mass, such as high pressure enclosure tubes or J-Young tubes, take extra care in seating the sample in depth gauge before inserting it in the magnet. If you see the missing indicator, you can use […]
Posted in Hardware, Spectrometer | No Comments
DRX499 – UPDATE : SYSTEM USABLE AGAIN
Apologies for the delay but the hung process has been cleared and users can start logging in to the instrument as before. NMR Facility Manager 10/15/15
Posted in Spectrometer, Usage | No Comments
UPDATE ! AV301 – BACK IN ACTION…
With the help of our ever dependable electronics shop, the TXD probe is fixed in house and we have AV301 system back on track with the same probehead. Users can engage the instrument as before.
Posted in Uncategorized | No Comments
UPDATE : AV301 : PROBE DAMAGED
The 1H tuning capacitor seems to have been forced beyond its limits and the probehead is unusable currently. We are working on a replacement probehead to take its place. The system will not be ready until tomorrow noon. Please watch this space for further updates.
Posted in Hardware | No Comments
ALERT: av301 TUNE & MATCH ISSUE
We are currently looking into an issue with the 1H tune and match on the av301. We will update when the issue is resolved.
Posted in Hardware, Spectrometer | No Comments