ALERT ! AV700 USERS : SYSTEM WILL BE DOWN FOR VT SYSTEM TROUBLESHOOT

Please wait further updates as to when the system will be back on line. NMR Facility Manager 6/12/2018

Posted in Hardware, Spectrometer | No Comments

UPDATE ! AV301 : SYSTEM BEING MONITORED…

We are aware of the unpredictable manner in which the acquisition gets interrupted once in a while. Thank you for the users who are engaging the instrument notwithstanding this condition and if you are feeling frustrated, I can totally understand.  Rest assured that we are closely collecting the OS level information when you engage the […]

Posted in Host Computer, Operating System | No Comments

UPDATE ! AV301 : SYSTEM IS RELEASED FOR USE – WITH A CAVEAT

There are two problems we are trying to troubleshoot that are of a sporadic nature.  As I write this, when a user logs in, everything might proceed as normal (including the UDRIVE mount).   It is also possible that the acquisition ends with an error message. In either case, the best recourse to pin down the […]

Posted in Host Computer, Operating System | No Comments

ALERT ! AV301 : DOWN UNTIL FURTHER NOTICE

From the diagnosis so far, it is a computer related glitch that is causing intermittent acquisition failure of the kind noticed by some of you. We are working to resolve this asap.  Please look for further updates. NMR Facility Manager 6/6/2018

Posted in Host Computer, Operating System | No Comments

UPDATE ! AV301 – SYSTEM BACK TO NORMAL

A software upgrade based bug caused the interruption today and we reverted the system to its earlier state, while troubleshooting the cause of the problem encountered.  That process will be continued independently on a different system.  As of now, the AV301 is behaving normally and the UDRIVE mounting is restored to normal Data written today […]

Posted in Datasets, Host Computer, Operating System | No Comments

ALERT ! AV301 – SYSTEM DOWN DUE TO COMPUTER ISSUE

We are working actively to resolve the issue and will post an update once the task is competed.  Other than the ‘udrive’ mounting problem, we are also aware of the acquisition error messages that was noticed by users. Hopefully, both these error conditions will be resolved at the same time. NMR Facility Manager 6/5/2018

Posted in Hardware, Spectrometer | No Comments

AV700 – UPDATE ON VT SYSTEM STATUS

As part of an ongoing investigation of the ultra low temperature chiller system, users are alerted to the following.  You will notice a Liquid Nitrogen Dewar sitting next to the Aluminum staircase. The N2 gas derived from this tank is used currently to run the VT system.  This setup is transparent to the way VT […]

Posted in Spectrometer, Usage | No Comments

UPDATE : AV700 USERS – SYSTEM READY FOR USE

The BBO-1H{X}-Z gradient probe is in place now. Please Remember: After creating a dataset, do ‘edsp’ to re-initialize the new base frequency. Load the default shim file : stdbbo and then start the shimming process. NMR Facility Manager 6/4/2018

Posted in Spectrometer, Usage | No Comments

ALERT ! AV700 USERS : IMMINENT PROBE CHANGE – Monday, 4 June ’18

The current 1H sensitive TXI-HCN probe will give way to the BBO-Z-gradient probe.  This probe is optimized for X-nuclei but also capable of 1H observe. The probe swap will occur on Monday the 4th of June 2018 forenoon. After the preliminary checks, the system should be ready for regular use by noon. Please contact me […]

Posted in Hardware, Spectrometer | No Comments

UPDATE ! AV700 – SYSTEM BACK IN OPERATION – NEW BASE FREQUENCY ! PLEASE READ

The Lock system is back to normal. PLEASE NOTE THE FOLLOWING IMPORTANT CHANGE ! New System Base Frequency :  699.962448 MHz. This reconfiguration is needed to compensate for the long term loss of energy by the Supercon Magnet. How does this affect you ? If you simply open your existing data set and start copying […]

Posted in Software, Spectrometer, Usage | No Comments