EARLY ALERT : POWER SHUTDOWN NEXT WEEK – Friday, Sep. 5th

Users might have noticed the power disruption announced for Bagley Hall on Sep. 6th, Saturday.  This planned event will affect the NMR lab Rm:37 from the early morning of Saturday 6th until about 3.00 pm on the same day.   The facility instruments, therefore will be shutdown by the afternoon of Friday the 5th Sep.  and […]

Posted in Hardware, Spectrometer | No Comments

AV 700 UPDATE: Returned to use with TXI probe

The problems with the av700 pneumatics board have been resolved. The AV700 is available for use with the solution TXI probe. The av700 will be changing to solids in a couple weeks.

Posted in Spectrometer, Usage | No Comments

DRX499 : UPDATE – 19F OBSERVE FULLY FUNCTIONAL

After creating a dataset of choice, users can simply enter the macro fluorine  to invoke the default 19F parameter set that includes  1H decoupling. Please remember to adjust O1P to your required offset to observe 19F signal.    Included below is a sample 19F spectrum.  Blue spectrum is with 1H decoupling and the Red spectrum is without 1H […]

Posted in Uncategorized | No Comments

AV500 – UPDATE : EVERYTHING LOOKS GOOD

System is tested to work fine.  Here is a solvent suppression snapshot done on the dummy sample i.e. 1:1 H2O+D2O.

Posted in Uncategorized | No Comments

AV700 – UPDATE : PNEUMATICS BOARD MALFUNCTION

Probably due to a voltage surge this weekend past,  the pneumatics control board on 700 appears compromised. No VT regulation or sample transport is possible due to this and we are energetically trying to resolve the issue.  Please look for further updates in this space. The machine is unusable until then.

Posted in Hardware, Spectrometer | No Comments

DRX499 : UPDATE – OPERATIONAL WITH LIMITED FUNCTIONALITY

The spectrometer with the TXO probe is operational fully with 1H and 13C channels. The 19F is not yet ready for prime use as I write this and I will post an update once that issue is resolved as well. The instrument is open for regular use, with the above caveat.

Posted in Hardware, Spectrometer, Usage | No Comments

UPDATE: AV300, AV301, AND AV500 AVAILABLE FOR USE

The errors being seen on the av300, av301, and av500 from the power fluctuations yesterday have been resolved. The spectrometers are now available for use. Please contact the facility managers and use the logbooks should you find further problems. Thanks.

Posted in Usage | No Comments

Errors in ‘rga’ on av300, av300, and av500

It appears the storms this morning caused some fluctuations in the electrical and gas lines to the NMR spectometers leading to issues with the consoles, as is being reported to the facility by users using the error logbooks currently. For the time being, please consider the NMR spectrometers offline until an NMR Facility Manager can […]

Posted in Uncategorized | No Comments

ALERT ! : PROBE CHANGE – AV500 and DRX499

Probe Swap date :  August 1st, Friday AV500 : The existing TXO probe will be replaced by the Z-gradient BBI probe and primed for use.  The system should be ready for use before the day is over. DRX499 : The TXO-19F probe will take the place of the current probe. Due to unavailability of gradient […]

Posted in Hardware, Spectrometer | 1 Comment

UPDATE – DRX499 : SYSTEM AVAILABLE – NEW COMPUTER UPGRADE INCOMPLETE

The new computer upgrade ran into problems with networking issues.  The old computer box is put in place and the spectrometer can continue to work as usual. The new computer plus new operating system will be troubleshot elsewhere and will be inducted into the spectrometer after all tests pass.   Sorry for the interruption and thank […]

Posted in Host Computer, Operating System | No Comments